The LUX-ZEPLIN (LZ) experiment will search for dark matter particle
interactions with a detector containing a total of 10 tonnes of liquid xenon
within a double-vessel cryostat. The large mass and proximity of the cryostat
to the active detector volume demand the use of material with extremely low
intrinsic radioactivity. We report on the radioassay campaign conducted to
identify suitable metals, the determination of factors limiting radiopure
production, and the selection of titanium for construction of the LZ cryostat
and other detector components. This titanium has been measured with activities
of $^{238}$U$_{e}$~$<$1.6~mBq/kg, $^{238}$U$_{l}$~$<$0.09~mBq/kg,
$^{232}$Th$_{e}$~$=0.28\pm 0.03$~mBq/kg, $^{232}$Th$_{l}$~$=0.25\pm
0.02$~mBq/kg, $^{40}$K~$<$0.54~mBq/kg, and $^{60}$Co~$<$0.02~mBq/kg (68\% CL).
Such low intrinsic activities, which are some of the lowest ever reported for
titanium, enable its use for future dark matter and other rare event searches.
Monte Carlo simulations have been performed to assess the expected background
contribution from the LZ cryostat with this radioactivity. In 1,000 days of
WIMP search exposure of a 5.6-tonne fiducial mass, the cryostat will contribute
only a mean background of $0.160\pm0.001$(stat)$\pm0.030$(sys) counts.