• A scintillating fiber array measurement system for gross beta is developed to achieve real-time monitoring of radioactivity in drinking water. The detector consists of 1,096 scintillating fibers, both sides of the fibers connect to a photomultiplier tube, and they are placed in a stainless steel tank. The detector parameters of working voltage, background counting rate and stability of the detector were tested, and the detection efficiency was calibrated using a standard solution of potassium chloride. Through experiment, the background counting rate of the detector is 38.31 cps and the detection efficiency for $\beta$ particles is 0.37 cps per Bq per liter; the detector can reach its detection limit of 1.0 Bq per liter for $\beta$ particles within 100 minutes without pre-concentration.
  • A prototype of LaBr3:Ce in situ gamma-ray spectrometer for marine environmental monitoring is developed and applied for in situ measurement. A 3-inch LaBr3:Ce scintillator is used in the detector, and a digital pulse process electronics is chosen as the pulse height analyzer. For this prototype, the energy response of the spectrometer is linear and the energy resolution of 662keV is 2.6% (much better than NaI). With the measurement of the prototype in a water tank filled with 137Cs, the detect efficiency for 137Cs is (0.288 0.01)cps/(Bq/L), which is close to the result of Monte Carlo simulation, 0.283cps/(Bq/L). With this measurement, the MDAC for 137Cs in one hour has been calculated to 0.78Bq/L, better than that of NaI(Tl) in-situ gamma spectrometer, which is ~1.0Bq/L.