In order to measure the energy of neutron fields, with energy ranging from 8
keV to 1 MeV, a new primary standard is being developed at the IRSN (Institute
for Radioprotection and Nuclear Safety). This project, micro-TPC (Micro Time
Projection Chamber), carried out in collaboration with the LPSC (Laboratoire de
Physique Subatomique et de Cosmologie), is based on the nuclear recoil detector
principle. The instrument is presented with the associated method to measure
the neutron energy. This article emphasizes the proton energy calibration
procedure and energy measurements of a neutron field produced at 127 keV with
the IRSN facility AMANDE.
-
Directional detection of non-baryonic Dark Matter is a promising search
strategy for discriminating WIMP events from neutrons, the ultimate background
for dark matter direct detection. This strategy requires both a precise
measurement of the energy down to a few keV and 3D reconstruction of tracks
down to a few mm. The MIMAC (MIcro-tpc MAtrix of Chambers) collaboration has
developed in the last years an original prototype detector based on the direct
coupling of large pixelized micromegas with a special developed fast
self-triggered electronics showing the feasibility of a new generation of
directional detectors. The first bi-chamber prototype has been installed at
Modane, underground laboratory in June 2012. The first undergournd background
events, the gain stability and calibration are shown. The first spectrum of
nuclear recoils showing 3D tracks coming from the radon progeny is presented.
-
The dark matter directional detection opens a new field in cosmology bringing
the possibility to build a map of nuclear recoils that would be able to explore
the galactic dark matter halo giving access to a particle characterization of
such matter and the shape of the halo. The MIMAC (MIcro-tpc MAtrix of Chambers)
collaboration has developed in the last years an original prototype detector
based on the direct coupling of large pixelized micromegas with a devoted fast
self-triggered electronics showing the feasibility of a new generation of
directional detectors. The discovery potential of this search strategy is
discussed and illustrated. In June 2012, the first bi-chamber prototype has
been installed at Modane Underground Laboratory (LSM) and the first underground
background events, the gain stability and calibration are shown.
-
A complete dedicated electronics, from front-end to back-end, was developed
to instrument a MIMAC prototype. A front end ASIC able to monitor 64 strips of
pixels and to provide their individual "Time Over Threshold" information has
been designed. An associated acquisition electronics and a real time track
reconstruction software have been developed to monitor a 512 channel prototype.
This auto-triggered electronic uses embedded processing to reduce the data
transfer to its useful part only, i.e. decoded coordinates of hit tracks and
corresponding energy measurements. The electronic designs, acquisition software
and the results obtained are presented.
-
The aim of the MIMAC project is to detect non-baryonic Dark Matter with a
directional TPC. The recent Micromegas efforts towards building a large size
detector will be described, in particular the characterization measurements of
a prototype detector of 10 $\times$ 10 cm$^2$ with a 2 dimensional readout
plane. Track reconstruction with alpha particles will be shown.
-
The aim of the MIMAC project is to detect non-baryonic Dark Matter with a
directional TPC using a high precision Micromegas readout plane. We will
describe in detail the recent developments done with bulk Micromegas detectors
as well as the characterisation measurements performed in an
Argon(95%)-Isobutane(5%) mixture. Track measurements with alpha particles will
be shown.
-
Directional detection of non-baryonic Dark Matter is a promising search
strategy for discriminating WIMP events from background. However, this strategy
requires both a precise measurement of the energy down to a few keV and 3D
reconstruction of tracks down to a few mm. To achieve this goal, the MIMAC
project has been developed. It is based on a gaseous micro-TPC matrix, filled
with 3He, CF4 and/or C4H10. The first results on low energy nuclear recoils (1H
and 19F) obtained with a low mono-energetic neutron field are presented. The
discovery potential of this search strategy is discussed and illustrated by a
realistic case accessible to MIMAC.
-
Directional detection of galactic Dark Matter is a promising search strategy
for discriminating geniune WIMP events from background ones. We present
technical progress on gaseous detectors as well as recent phenomenological
studies, allowing the design and construction of competitive experiments.