The scientific performance of the Planck Low Frequency Instrument (LFI) after
one year of in-orbit operation is presented. We describe the main optical
parameters and discuss photometric calibration, white noise sensitivity, and
noise properties. A preliminary evaluation of the impact of the main systematic
effects is presented. For each of the performance parameters, we outline the
methods used to obtain them from the flight data and provide a comparison with
pre-launch ground assessments, which are essentially confirmed in flight.
-
The Low Frequency Instrument (LFI) on-board the ESA Planck satellite carries
eleven radiometer subsystems, called Radiometer Chain Assemblies (RCAs), each
composed of a pair of pseudo-correlation receivers. We describe the on-ground
calibration campaign performed to qualify the flight model RCAs and to measure
their pre-launch performances. Each RCA was calibrated in a dedicated
flight-like cryogenic environment with the radiometer front-end cooled to 20K
and the back-end at 300K, and with an external input load cooled to 4K. A
matched load simulating a blackbody at different temperatures was placed in
front of the sky horn to derive basic radiometer properties such as noise
temperature, gain, and noise performance, e.g. 1/f noise. The spectral response
of each detector was measured as was their susceptibility to thermal variation.
All eleven LFI RCAs were calibrated. Instrumental parameters measured in these
tests, such as noise temperature, bandwidth, radiometer isolation, and
linearity, provide essential inputs to the Planck-LFI data analysis.
-
The Planck Low Frequency Instrument (LFI) radiometers have been tested
extensively during several dedicated campaigns. The present paper reports the
principal noise properties of the LFI radiometers.
-
In this paper we discuss the linearity response of the Planck-LFI receivers,
with particular reference to signal compression measured on the 30 and 44 GHz
channels. In the article we discuss the various sources of compression and
present a model that accurately describes data measured during tests performed
with individual radiomeric chains. After discussing test results we present the
best parameter set representing the receiver response and discuss the impact of
non linearity on in-flight calibration, which is shown to be negligible.
-
This paper is part of the Prelaunch status LFI papers published on JINST:
http://www.iop.org/EJ/journal/-page=extra.proc5/jinst
The Planck LFI Radiometer Chain Assemblies (RCAs) have been calibrated in two
dedicated cryogenic facilities. In this paper the facilities and the related
instrumentation are described. The main satellite thermal interfaces for the
single chains have to be reproduced and stability requirements have to be
satisfied. Setup design, problems occurred and improving solutions implemented
are discussed. Performance of the cryogenic setup are reported.
-
The Low Frequency Instrument (LFI) is a radiometer array covering the 30-70
GHz spectral range on-board the ESA Planck satellite, launched on May 14th,
2009 to observe the cosmic microwave background (CMB) with unprecedented
precision. In this paper we describe the development and validation of a
software model of the LFI pseudo-correlation receivers which enables to
reproduce and predict all the main system parameters of interest as measured at
each of the 44 LFI detectors. These include system total gain, noise
temperature, band-pass response, non-linear response. The LFI Advanced RF Model
(LARFM) has been constructed by using commercial software tools and data of
each radiometer component as measured at single unit level. The LARFM has been
successfully used to reproduce the LFI behavior observed during the LFI
ground-test campaign. The model is an essential element in the database of LFI
data processing center and will be available for any detailed study of
radiometer behaviour during the survey.
-
We give the calibration and scientific performance parameters of the Planck
Low Frequency Instrument (LFI) measured during the ground cryogenic test
campaign. These parameters characterise the instrument response and constitute
our best pre-launch knowledge of the LFI scientific performance. The LFI shows
excellent $1/f$ stability and rejection of instrumental systematic effects;
measured noise performance shows that LFI is the most sensitive instrument of
its kind. The set of measured calibration parameters will be updated during
flight operations through the end of the mission.
-
In this paper we present the Low Frequency Instrument (LFI), designed and
developed as part of the Planck space mission, the ESA program dedicated to
precision imaging of the cosmic microwave background (CMB). Planck-LFI will
observe the full sky in intensity and polarisation in three frequency bands
centred at 30, 44 and 70 GHz, while higher frequencies (100-850 GHz) will be
covered by the HFI instrument. The LFI is an array of microwave radiometers
based on state-of-the-art Indium Phosphide cryogenic HEMT amplifiers
implemented in a differential system using blackbody loads as reference
signals. The front-end is cooled to 20K for optimal sensitivity and the
reference loads are cooled to 4K to minimise low frequency noise. We provide an
overview of the LFI, discuss the leading scientific requirements and describe
the design solutions adopted for the various hardware subsystems. The main
drivers of the radiometric, optical and thermal design are discussed, including
the stringent requirements on sensitivity, stability, and rejection of
systematic effects. Further details on the key instrument units and the results
of ground calibration are provided in a set of companion papers.