We studied the application of statistical reconstruction algorithms, namely
maximum likelihood and least squares methods, to the problem of event
reconstruction in a dual phase liquid xenon detector. An iterative method was
developed for in-situ reconstruction of the PMT light response functions from
calibration data taken with an uncollimated gamma-ray source. Using the
techniques described, the performance of the ZEPLIN-III dark matter detector
was studied for 122 keV gamma-rays. For the inner part of the detector (R<100
mm), spatial resolutions of 13 mm and 1.6 mm FWHM were measured in the
horizontal plane for primary and secondary scintillation, respectively. An
energy resolution of 8.1% FWHM was achieved at that energy. The possibility of
using this technique for improving performance and reducing cost of
scintillation cameras for medical applications is currently under study.
-
We study a three stage dark matter and neutrino observatory based on
multi-ton two-phase liquid Xe and Ar detectors with sufficiently low
backgrounds to be sensitive to WIMP dark matter interaction cross sections down
to 10E-47 cm^2, and to provide both identification and two independent
measurements of the WIMP mass through the use of the two target elements in a
5:1 mass ratio, giving an expected similarity of event numbers. The same
detection systems will also allow measurement of the pp solar neutrino
spectrum, the neutrino flux and temperature from a Galactic supernova, and
neutrinoless double beta decay of 136Xe to the lifetime level of 10E27 - 10E28
y corresponding to the Majorana mass predicted from current neutrino
oscillation data. The proposed scheme would be operated in three stages G2, G3,
G4, beginning with fiducial masses 1-ton Xe + 5-ton Ar (G2), progressing to
10-ton Xe + 50-ton Ar (G3) then, dependent on results and performance of the
latter, expandable to 100-ton Xe + 500-ton Ar (G4). This method of scale-up
offers the advantage of utilizing the Ar vessel and ancillary systems of one
stage for the Xe detector of the succeeding stage, requiring only one new
detector vessel at each stage. Simulations show the feasibility of reducing or
rejecting all external and internal background levels to a level <1 events per
year for each succeeding mass level, by utilizing an increasing outer thickness
of target material as self-shielding. The system would, with increasing mass
scale, become increasingly sensitive to annual signal modulation, the agreement
of Xe and Ar results confirming the Galactic origin of the signal. Dark matter
sensitivities for spin-dependent and inelastic interactions are also included,
and we conclude with a discussion of possible further gains from the use of
Xe/Ar mixtures.
-
Plastic scintillators are widely used in industry, medicine and scientific
research, including nuclear and particle physics. Although one of their most
common applications is in neutron detection, experimental data on their
response to low-energy nuclear recoils are scarce. Here, the relative
scintillation efficiency for neutron-induced nuclear recoils in a
polystyrene-based plastic scintillator (UPS-923A) is presented, exploring
recoil energies between 125 keV and 850 keV. Monte Carlo simulations,
incorporating light collection efficiency and energy resolution effects, are
used to generate neutron scattering spectra which are matched to observed
distributions of scintillation signals to parameterise the energy-dependent
quenching factor. At energies above 300 keV the dependence is reasonably
described using the semi-empirical formulation of Birks and a kB factor of
(0.014+/-0.002) g/MeVcm^2 has been determined. Below that energy the measured
quenching factor falls more steeply than predicted by the Birks formalism.
-
Scintillation and ionisation yields for nuclear recoils in liquid xenon above
10 keVnr (nuclear recoil energy) are deduced from data acquired using broadband
Am-Be neutron sources. The nuclear recoil data from several exposures to two
sources were compared to detailed simulations. Energy-dependent scintillation
and ionisation yields giving acceptable fits to the data were derived.
Efficiency and resolution effects are treated using a light collection Monte
Carlo, measured photomultiplier response profiles and hardware trigger studies.
A gradual fall in scintillation yield below ~40 keVnr is found, together with a
rising ionisation yield; both are in good agreement with the latest independent
measurements. The analysis method is applied to both the most recent ZEPLIN-III
data, acquired with a significantly upgraded detector and a
precision-calibrated Am-Be source, as well as to the earlier data from the
first run in 2008. A new method for deriving the recoil scintillation yield,
which includes sub-threshold S1 events, is also presented which confirms the
main analysis.
-
We present an experimental study of single electron emission in ZEPLIN-III, a
two-phase xenon experiment built to search for dark matter WIMPs, and discuss
applications enabled by the excellent signal-to-noise ratio achieved in
detecting this signature. Firstly, we demonstrate a practical method for
precise measurement of the free electron lifetime in liquid xenon during normal
operation of these detectors. Then, using a realistic detector response model
and backgrounds, we assess the feasibility of deploying such an instrument for
measuring coherent neutrino-nucleus elastic scattering using the ionisation
channel in the few-electron regime. We conclude that it should be possible to
measure this elusive neutrino signature above an ionisation threshold of
$\sim$3 electrons both at a stopped pion source and at a nuclear reactor.
Detectable signal rates are larger in the reactor case, but the triggered
measurement and harder recoil energy spectrum afforded by the accelerator
source enable lower overall background and fiducialisation of the active
volume.
-
We examine electron and nuclear recoil backgrounds from radioactivity in the
ZEPLIN-III dark matter experiment at Boulby. The rate of low-energy electron
recoils in the liquid xenon WIMP target is 0.75$\pm$0.05 events/kg/day/keV,
which represents a 20-fold improvement over the rate observed during the first
science run. Energy and spatial distributions agree with those predicted by
component-level Monte Carlo simulations propagating the effects of the
radiological contamination measured for materials employed in the experiment.
Neutron elastic scattering is predicted to yield 3.05$\pm$0.5 nuclear recoils
with energy 5-50 keV per year, which translates to an expectation of 0.4 events
in a 1-year dataset in anti-coincidence with the veto detector for realistic
signal acceptance. Less obvious background sources are discussed, especially in
the context of future experiments. These include contamination of scintillation
pulses with Cherenkov light from Compton electrons and from $\beta$ activity
internal to photomultipliers, which can increase the size and lower the
apparent time constant of the scintillation response. Another challenge is
posed by multiple-scatter $\gamma$-rays with one or more vertices in regions
that yield no ionisation. If the discrimination power achieved in the first run
can be replicated, ZEPLIN-III should reach a sensitivity of $\sim 1 \times
10^{-8}$ pb$\cdot$year to the scalar WIMP-nucleon elastic cross-section, as
originally conceived.
-
The design, optimisation and construction of an anti-coincidence veto
detector to complement the ZEPLIN-III direct dark matter search instrument is
described. One tonne of plastic scintillator is arranged into 52 bars
individually read out by photomultipliers and coupled to a gadolinium-loaded
passive polypropylene shield. Particular attention has been paid to
radiological content. The overall aim has been to achieve a veto detector of
low threshold and high efficiency without the creation of additional background
in ZEPLIN-III, all at a reasonable cost. Extensive experimental measurements of
the components have been made, including radioactivity levels and performance
characteristics. These have been used to inform a complete end-to-end Monte
Carlo simulation that has then been used to calculate the expected performance
of the new instrument, both operating alone and as an anti-coincidence detector
for ZEPLIN-III. The veto device will be capable of rejecting over 65% of
coincident nuclear recoil events from neutron background in the energy range of
interest in ZEPLIN-III. This will reduce the background in ZEPLIN-III from ~0.4
to ~0.14 events per year in the WIMP acceptance region, a significant factor in
the event of a non-zero observation. Furthermore, in addition to providing
valuable diagnostic capabilities, the veto is capable of tagging over 15% for
gamma-ray rejection, all whilst contributing no significant additional
background. In conjunction with the replacement of the internal ZEPLIN-III
photomultiplier array, the new veto is expected to improve significantly the
sensitivity of the ZEPLIN-III instrument to dark matter, allowing spin
independent WIMP-nucleon cross sections below 1E-8 pb to be probed.
-
We present limits on the WIMP-nucleon cross section for inelastic dark matter
derived from the 2008 run of ZEPLIN-III. Cuts, notably on scintillation pulse
shape and scintillation-to-ionisation ratio, give a net exposure of 63 kg.days
in the range 20-80keV nuclear recoil energy, in which 6 events are observed.
Upper limits on signal rate are derived from the maximum empty patch in the
data. Under standard halo assumptions a small region of parameter space
consistent, at 99% CL, with causing the 1.17 ton.year DAMA modulation signal is
allowed at 90% CL: it is in the mass range 45-60 GeV with a minimum CL of 88%,
again derived from the maximum patch. This is the tightest constraint on that
explanation of the DAMA result yet presented using a xenon target.
-
The ZEPLIN-III experiment in the Palmer Underground Laboratory at Boulby uses
a 12kg two-phase xenon time projection chamber to search for the weakly
interacting massive particles (WIMPs) that may account for the dark matter of
our Galaxy. The detector measures both scintillation and ionisation produced by
radiation interacting in the liquid to differentiate between the nuclear
recoils expected from WIMPs and the electron recoil background signals down to
~10keV nuclear recoil energy. An analysis of 847kg.days of data acquired
between February 27th 2008 and May 20th 2008 has excluded a WIMP-nucleon
elastic scattering spin-independent cross-section above 8.1x10(-8)pb at
55GeV/c2 with a 90% confidence limit. It has also demonstrated that the
two-phase xenon technique is capable of better discrimination between electron
and nuclear recoils at low-energy than previously achieved by other xenon-based
experiments.
-
A method is described that allows calibration and assessment of the linearity
of response of an array of photomultiplier tubes. The method does not require
knowledge of the photomultiplier single photoelectron response model and uses
science data directly, thus eliminating the need for dedicated data sets. In
this manner all photomultiplier working conditions (e.g. temperature, external
fields, etc) are exactly matched between calibration and science acquisitions.
This is of particular importance in low background experiments such as
ZEPLIN-III, where methods involving the use of external light sources for
calibration are severely constrained.
-
We present new experimental constraints on the WIMP-nucleon spin-dependent
elastic cross-sections using data from the first science run of ZEPLIN-III, a
two-phase xenon experiment searching for galactic dark matter WIMPs based at
the Boulby mine. Analysis of $\sim$450 kg$\cdot$days fiducial exposure revealed
a most likely signal of zero events, leading to a 90%-confidence upper limit on
the pure WIMP-neutron cross-section of $\sigma_n=1.8\times 10^{-2}$ pb at 55
GeV/$c^2$ WIMP mass. Recent calculations of the nuclear spin structure based on
the Bonn CD nucleon-nucleon potential were used for the odd-neutron isotopes
$^{129}$Xe and $^{131}$Xe. These indicate that the sensitivity of xenon targets
to the spin-dependent WIMP-proton interaction is much lower than implied by
previous calculations, whereas the WIMP-neutron sensitivity is impaired only by
a factor of $\sim$2.
-
A multi-target detection system XAX, comprising concentric 10 ton targets of
136Xe and 129/131Xe, together with a geometrically similar or larger target of
liquid Ar, is described. Each is configured as a two-phase
scintillation/ionization TPC detector, enhanced by a full 4pi array of
ultra-low radioactivity Quartz Photon Intensifying Detectors (QUPIDs) replacing
the conventional photomultipliers for detection of scintillation light. It is
shown that background levels in XAX can be reduced to the level required for
dark matter particle (WIMP) mass measurement at a 10^-10 pb WIMP-nucleon cross
section, with single-event sensitivity below 10^-11 pb. The use of multiple
target elements allows for confirmation of the A^2 dependence of a coherent
cross section, and the different Xe isotopes provide information on the
spin-dependence of the dark matter interaction. The event rates observed by Xe
and Ar would modulate annually with opposite phases from each other for WIMP
mass >~100 GeV/c^2. The large target mass of 136Xe and high degree of
background reduction allow neutrinoless double beta decay to be observed with
lifetimes of 10^27-10^28 years, corresponding to the Majorana neutrino mass
range 0.01-0.1 eV, the most likely range from observed neutrino mass
differences. The use of a 136Xe-depleted 129/131Xe target will also allow
measurement of the pp solar neutrino spectrum to a precision of 1-2%.
-
We present the first measurements of the muon-induced neutron flux at the
Boulby Underground Laboratory. The experiment was carried out with an 0.73
tonne liquid scintillator that also served as an anticoincidence system for the
ZEPLIN-II direct dark matter search. The experimental method exploited the
delayed coincidences between high-energy muon signals and gamma-rays from
radiative neutron capture on hydrogen or other elements. The muon-induced
neutron rate, defined as the average number of detected neutrons per detected
muon, was measured as $0.079 \pm 0.003$ (stat.) neutrons/muon using
neutron-capture signals above 0.55 MeV in a time window of 40-190 $\mu$s after
the muon trigger. Accurate Monte Carlo simulations of the neutron production,
transport and detection in a precisely modeled laboratory and experimental
setup using the GEANT4 toolkit gave a result 1.8 times higher than the measured
value. The difference greatly exceeds all statistical and systematic
uncertainties. As the vast majority of neutrons detected in the current setup
were produced in lead we evaluated from our measurements the neutron yield in
lead as $(1.31 \pm 0.06) \times 10^{-3}$ neutrons/muon/(g/cm$^2$) for a mean
muon energy of about 260 GeV.
-
We present the first measurements of the electroluminescence response to the
emission of single electrons in a two-phase noble gas detector. Single
ionization electrons generated in liquid xenon are detected in a thin gas layer
during the 31-day background run of the ZEPLIN-II experiment, a two-phase xenon
detector for WIMP dark matter searches. Both the pressure dependence and
magnitude of the single-electron response are in agreement with previous
measurements of electroluminescence yield in xenon. We discuss different
photoionization processes as possible cause for the sample of single electrons
studied in this work. This observation may have implications for the design and
operation of future large-scale two-phase systems.
-
We propose a new concept for a very low background multi-ton liquid xenon
Dark Matter experiment. The detector consists of two concentric spheres and a
charge readout device in the centre. Xenon between the two spheres forms a
self-shield and veto device. The inner surface of the central sphere is coated
with CsI to form an internal photocathode with minimum of 2\pi coverage for any
event in the active volume. Photoelectrons from the CsI photocathode drift
toward the charge readout micro-structure in the centre of the detector. Both
scintillation and ionisation are measured simultaneously for background
rejection and 3-D event mapping. In addition to external shielding, the low
background is achieved by eliminating PMTs and by using low radioactivity pure
materials throughout the detector. We present detailed calculations of the
charge readout system and design details. The detector is expected to probe the
full SUSY parameter space.
-
Results are presented from the first underground data run of ZEPLIN-II, a 31
kg two phase xenon detector developed to observe nuclear recoils from
hypothetical weakly interacting massive dark matter particles. Discrimination
between nuclear recoils and background electron recoils is afforded by
recording both the scintillation and ionisation signals generated within the
liquid xenon, with the ratio of these signals being different for the two
classes of event. This ratio is calibrated for different incident species using
an AmBe neutron source and Co-60 gamma-ray sources. From our first 31 live days
of running ZEPLIN-II, the total exposure following the application of fiducial
and stability cuts was 225 kgxdays. A background population of radon progeny
events was observed in this run, arising from radon emission in the gas
purification getters, due to radon daughter ion decays on the surfaces of the
walls of the chamber. An acceptance window, defined by the neutron calibration
data, of 50% nuclear recoil acceptance between 5 keVee and 20 keVee, had an
observed count of 29 events, with a summed expectation of 28.6+/-4.3 gamma-ray
and radon progeny induced background events. These figures provide a 90% c.l.
upper limit to the number of nuclear recoils of 10.4 events in this acceptance
window, which converts to a WIMP-nucleon spin-independent cross-section with a
minimum of 6.6x10^-7 pb following the inclusion of an energy dependent,
calibrated, efficiency. A second run is currently underway in which the radon
progeny will be eliminated, thereby removing the background population, with a
projected sensitivity of 2x10^-7 pb for similar exposures as the first run.
-
We present details of the technical design and manufacture of the ZEPLIN-III
dark matter experiment. ZEPLIN-III is a two-phase xenon detector which measures
both the scintillation light and the ionisation charge generated in the liquid
by interacting particles and radiation. The instrument design is driven by both
the physics requirements and by the technology requirements surrounding the use
of liquid xenon. These include considerations of key performance parameters,
such as the efficiency of scintillation light collection, restrictions placed
on the use of materials to control the inherent radioactivity levels,
attainment of high vacuum levels and chemical contamination control. The
successful solution has involved a number of novel design and manufacturing
features which will be of specific use to future generations of direct dark
matter search experiments as they struggle with similar and progressively more
demanding requirements.
-
We present results from a GEANT4-based Monte Carlo tool for end-to-end
simulations of the ZEPLIN-III dark matter experiment. ZEPLIN-III is a two-phase
detector which measures both the scintillation light and the ionisation charge
generated in liquid xenon by interacting particles and radiation. The software
models the instrument response to radioactive backgrounds and calibration
sources, including the generation, ray-tracing and detection of the primary and
secondary scintillations in liquid and gaseous xenon, and subsequent processing
by data acquisition electronics. A flexible user interface allows easy
modification of detector parameters at run time. Realistic datasets can be
produced to help with data analysis, an example of which is the position
reconstruction algorithm developed from simulated data. We present a range of
simulation results confirming the original design sensitivity of a few times
$10^{-8}$ pb to the WIMP-nucleon cross-section.
-
The neutron background at the underground laboratory at Boulby - a site for
several dark matter experiments - is discussed. Special emphasis is put on the
neutron background produced by cosmic-ray muons. The most recent versions of
the muon propagation code MUSIC, and particle transport code FLUKA are used to
evaluate muon and neutron fluxes. The results of simulations are compared with
experimental data.
-
The NAIAD experiment (NaI Advanced Detector) for WIMP dark matter searches at
Boulby mine (UK) is described. The detector consists of an array of
encapsulated and unencapsulated NaI(Tl) crystals with high light yield. Six
crystals are collecting data at present. Data accumulated by four of them (10.6
kg x year exposure) have been used to set upper limits on the WIMP-nucleon
spin-independent and WIMP-proton spin-dependent cross-sections. Pulse shape
analysis has been applied to discriminate between nuclear recoils, as may be
caused by WIMP interactions, and electron recoils due to gamma background.
Various calibrations of crystals are presented.
-
The status of dark matter searches with inorganic scintillator detectors at
Boulby mine is reviewed and the results of tests with a CsI(Tl) crystal are
presented. The objectives of the latter experiment were to study anomalous fast
events previously observed and to identify ways to remove this background.
Clear indications were found that these events were due to surface
contamination of crystals by alphas, probably from radon decay. A new array of
unencapsulated NaI(Tl) crystals immersed either in liquid paraffin or pure
nitrogen atmosphere is under construction at Boulby. Such an approach allows
complete control of the surface of the crystals and the ability to remove any
surface contamination. First data from the unencapsulated NaI(Tl) do not show
the presence of anomalous fast events.
-
Measurements of the shapes of scintillation pulses produced by nuclear
recoils, alpha particles and photons in NaI(Tl) crystals at visible energies of
10-100 keV have been performed in order to investigate possible sources of
background in NaI(Tl) dark matter experiments and, in particular, the possible
origin of the anomalous fast time constant events observed in the UK Dark
Matter Collaboration experiments at Boulby mine. Pulses initiated by X-rays
(via photoelectric effect close to the surface of the crystal) were found not
to differ from those produced by high-energy photons (via Compton electrons
inside the crystal) within experimental errors. However, pulses induced by
alpha particles (degraded from an external MeV source) were found to be ~10%
faster than those of nuclear recoils, but insufficiently fast to account for
the anomalous events.