The \MJ\ Collaboration is operating an array of high purity Ge detectors to
search for neutrinoless double-beta decay in $^{76}$Ge. The \MJ\ \DEM\
comprises 44.1~kg of Ge detectors (29.7 kg enriched in $^{76}$Ge) split between
two modules contained in a low background shield at the Sanford Underground
Research Facility in Lead, South Dakota. Here we present results from data
taken during construction, commissioning, and the start of full operations. We
achieve unprecedented energy resolution of 2.5 keV FWHM at \qval\ and a very
low background with no observed candidate events in 10 kg yr of enriched Ge
exposure, resulting in a lower limit on the half-life of $1.9\times10^{25}$ yr
(90\% CL). This result constrains the effective Majorana neutrino mass to below
240 to 520 meV, depending on the matrix elements used. In our experimental
configuration with the lowest background, the background is $4.0_{-2.5}^{+3.1}$
counts/(FWHM t yr).
-
The next generation low-background detectors operating deep underground aim
for unprecedented low levels of radioactive backgrounds. The deposition and
presence of radon progeny on detector surfaces is an added source of energetic
background events. In addition to limiting the detector material's radon
exposure in order to reduce potential surface backgrounds, it is just as
important to clean surfaces to remove inevitable contamination. Such studies of
radon progeny removal have generally found that a form of etching is effective
at removing some of the progeny (Bi and Pb), however more aggressive
techniques, including electropolishing, have been shown to effectively remove
the Po atoms. In the absence of an aggressive etch, a significant fraction of
the Po atoms are believed to either remain behind within the surface or
redeposit from the etching solution back onto the surface. We explore the
chemical nature of the aqueous Po ions and the effect of the oxidation state of
Po to maximize the Po ions remaining in the etching solution of contaminated Cu
surfaces. We present a review of the previous studies of surface radon progeny
removal and our findings on the role of oxidizing agents and a cell potential
in the preparation of a clean etching technique.
-
The MAJORANA Collaboration is searching for the neutrinoless double-beta
decay of the nucleus Ge-76. The MAJORANA DEMONSTRATOR is an array of germanium
detectors deployed with the aim of implementing background reduction techniques
suitable for a tonne scale Ge-76-based search (the LEGEND collaboration). In
the DEMONSTRATOR, germanium detectors operate in an ultra-pure vacuum cryostat
at 80 K. One special challenge of an ultra-pure environment is to develop
reliable cables, connectors, and electronics that do not significantly
contribute to the radioactive background of the experiment. This paper
highlights the experimental requirements and how these requirements were met
for the MAJORANA DEMONSTRATOR, including plans to upgrade the wiring for higher
reliability in the summer of 2018. Also described are requirements for LEGEND
R&D efforts underway to meet these additional requirements.
-
The MAJORANA DEMONSTRATOR is a neutrinoless double beta decay experiment
utilizing enriched Ge-76 detectors in 2 separate modules inside of a common
solid shield at the Sanford Underground Research Facility. The DEMONSTRATOR has
utilized world leading assay sensitivities to develop clean materials and
processes for producing ultra-pure copper and plastic components. This
experiment is now operating, and initial data provide new insights into the
success of cleaning and processing. Post production copper assays after the
completion of Module 1 showed an increase in U and Th contamination in finished
parts compared to starting bulk material. A revised cleaning method and
additional round of surface contamination studies prior to Module 2
construction have provided evidence that more rigorous process control can
reduce surface contamination. This article describes the assay results and
discuss further studies to take advantage of assay capabilities for the purpose
of maintaining ultra clean fabrication and process design.
-
The observation of neutrinoless double-beta decay (0${\nu}{\beta}{\beta}$)
would show that lepton number is violated, reveal that neutrinos are Majorana
particles, and provide information on neutrino mass. A discovery-capable
experiment covering the inverted ordering region, with effective Majorana
neutrino masses of 15 - 50 meV, will require a tonne-scale experiment with
excellent energy resolution and extremely low backgrounds, at the level of
$\sim$0.1 count /(FWHM$\cdot$t$\cdot$yr) in the region of the signal. The
current generation $^{76}$Ge experiments GERDA and the MAJORANA DEMONSTRATOR
utilizing high purity Germanium detectors with an intrinsic energy resolution
of 0.12%, have achieved the lowest backgrounds by over an order of magnitude in
the 0${\nu}{\beta}{\beta}$ signal region of all 0${\nu}{\beta}{\beta}$
experiments. Building on this success, the LEGEND collaboration has been formed
to pursue a tonne-scale $^{76}$Ge experiment. The collaboration aims to develop
a phased 0${\nu}{\beta}{\beta}$ experimental program with discovery potential
at a half-life approaching or at $10^{28}$ years, using existing resources as
appropriate to expedite physics results.
-
Neutrinoless double-beta decay searches play a major role in determining the
nature of neutrinos, the existence of a lepton violating process, and the
effective Majorana neutrino mass. The MAJORANA Collaboration assembled an array
of high purity Ge detectors to search for neutrinoless double-beta decay in
Ge-76. The MAJORANA DEMONSTRATOR is comprised of 44.1 kg (29.7 kg enriched in
Ge-76) of Ge detectors divided between two modules contained in a
low-background shield at the Sanford Underground Research Facility in Lead,
South Dakota, USA. The initial goals of the DEMONSTRATOR are to establish the
required background and scalability of a Ge-based next-generation ton-scale
experiment. Following a commissioning run that started in 2015, the first
detector module started low-background data production in early 2016. The
second detector module was added in August 2016 to begin operation of the
entire array. We discuss results of the initial physics runs, as well as the
status and physics reach of the full MAJORANA DEMONSTRATOR experiment.
-
The MAJORANA DEMONSTRATOR is an array of point-contact Ge detectors
fabricated from Ge isotopically enriched to 88% in Ge-76 to search for
neutrinoless double beta decay. The processing of Ge for germanium detectors is
a well-known technology. However, because of the high cost of Ge enriched in
Ge-76, special procedures were required to maximize the yield of detector mass
and to minimize exposure to cosmic rays. These procedures include careful
accounting for the material; shielding it to reduce cosmogenic generation of
radioactive isotopes; and development of special reprocessing techniques for
contaminated solid germanium, shavings, grindings, acid etchant and cutting
fluids from detector fabrication. Processing procedures were developed that
resulted in a total yield in detector mass of 70%. However, none of the
acid-etch solution and only 50% of the cutting fluids from detector fabrication
were reprocessed. Had they been processed, the projections for the recovery
yield would be between 80 -- 85%. Maximizing yield is critical to justify a
possible future ton-scale experiment. A process for recovery of germanium from
the acid-etch solution was developed with yield of about 90%. All material was
shielded or stored underground whenever possible to minimize the formation of
Ge-68 by cosmic rays, which contributes background in the double-beta decay
region of interest and cannot be removed by zone refinement and crystal growth.
Formation of Ge-68 was reduced by a significant factor over that in natural
abundance detectors not protected from cosmic rays.
-
We present new limits on exotic keV-scale physics based on 478 kg d of
MAJORANA DEMONSTRATOR commissioning data. Constraints at the 90% confidence
level are derived on bosonic dark matter (DM) and solar axion couplings, Pauli
exclusion principle violating (PEPV) decay, and electron decay using
monoenergetic peak signal-limits above our background. Our most stringent DM
constraints are set for 11.8 keV mass particles, limiting $g_{Ae} <4.5\times
10^{-13}$ for pseudoscalars and $\frac{\alpha'}{\alpha} < 9.7\times 10^{-28}$
for vectors. We also report a 14.4 keV solar axion coupling limit of
$g_{AN}^{\mathrm{eff}}\times g_{Ae}~<~3.8 \times 10^{-17}$, a
$\frac{1}{2}\beta^2~<~8.5\times10^{-48}$ limit on the strength of PEPV electron
transitions, and a lower limit on the electron lifetime of $\tau_e > 1.2 \times
10^{24}\;$yr for $e^- \rightarrow$ invisible.
-
The MAJORANA Collaboration is searching for the neutrinoless double-beta
decay of the nucleus $^{76}$Ge. The MAJORANA DEMONSTRATOR is an array of
germanium detectors deployed with the aim of implementing background reduction
techniques suitable for a 1-tonne $^{76}$Ge-based search. The ultra
low-background conditions require regular calibrations to verify proper
function of the detectors. Radioactive line sources can be deployed around the
cryostats containing the detectors for regular energy calibrations. When
measuring in low-background mode, these line sources have to be stored outside
the shielding so they do not contribute to the background. The deployment and
the retraction of the source are designed to be controlled by the data
acquisition system and do not require any direct human interaction. In this
paper, we detail the design requirements and implementation of the calibration
apparatus, which provides the event rates needed to define the pulse-shape cuts
and energy calibration used in the final analysis as well as data that can be
compared to simulations.
-
A search for Pauli-exclusion-principle-violating K-alpha electron transitions
was performed using 89.5 kg-d of data collected with a p-type point contact
high-purity germanium detector operated at the Kimballton Underground Research
Facility. A lower limit on the transition lifetime of 5.8x10^30 seconds at 90%
C.L. was set by looking for a peak at 10.6 keV resulting from the x-ray and
Auger electrons present following the transition. A similar analysis was done
to look for the decay of atomic K-shell electrons into neutrinos, resulting in
a lower limit of 6.8x10^30 seconds at 90 C.L. It is estimated that the MAJORANA
DEMONSTRATOR, a 44 kg array of p-type point contact detectors that will search
for the neutrinoless double-beta decay of 76-Ge, could improve upon these
exclusion limits by an order of magnitude after three years of operation.
-
The Majorana Demonstrator searches for neutrinoless double-beta decay of
$^{76}$Ge using arrays of high-purity germanium detectors. If observed, this
process would demonstrate that lepton number is not a conserved quantity in
nature, with implications for grand-unification and for explaining the
predominance of matter over antimatter in the universe. A problematic
background in such large granular detector arrays is posed by alpha particles.
In the Majorana Demonstrator, events have been observed that are consistent
with energy- degraded alphas originating on the passivated surface, leading to
a potential background contribution in the region-of-interest for neutrinoless
double-beta decay. However, it is also observed that when energy deposition
occurs very close to the passivated surface, charges drift through the bulk
onto that surface, and then drift along it with greatly reduced mobility. This
leads to both a reduced prompt signal and a measurable change in slope of the
tail of a recorded pulse. In this contribution we discuss the characteristics
of these events and the development of a filter that can identify the
occurrence of this delayed charge recovery, allowing for the efficient
rejection of passivated surface alpha events in analysis.
-
Neutrinoless double-beta decay searches seek to determine the nature of
neutrinos, the existence of a lepton violating process, and the effective
Majorana neutrino mass. The {\sc Majorana} Collaboration is assembling an array
of high purity Ge detectors to search for neutrinoless double-beta decay in
$^{76}$Ge. The {\sc Majorana Demonstrator} is composed of 44.8~kg (29.7 kg
enriched in $^{76}$Ge) of Ge detectors in total, split between two modules
contained in a low background shield at the Sanford Underground Research
Facility in Lead, South Dakota. The initial goals of the {\sc Demonstrator} are
to establish the required background and scalability of a Ge-based,
next-generation, tonne-scale experiment. Following a commissioning run that
began in 2015, the first detector module started physics data production in
early 2016. We will discuss initial results of the Module 1 commissioning and
first physics run, as well as the status and potential physics reach of the
full {\sc Majorana Demonstrator} experiment. The collaboration plans to
complete the assembly of the second detector module by mid-2016 to begin full
data production with the entire array.
-
The MAJORANA Collaboration is constructing a system containing 44 kg of
high-purity Ge (HPGe) detectors to demonstrate the feasibility and potential of
a future tonne-scale experiment capable of probing the neutrino mass scale to
~15 meV. To realize this, a major goal of the MAJORANA DEMONSTRATOR is to
demonstrate a path forward to achieving a background rate at or below 1
count/(ROI-t-y) in the 4 keV region of interest (ROI) around the Q-value at
2039 keV. This goal is pursued through a combination of a significant reduction
of radioactive impurities in construction materials and analytical methods for
background rejection, for example using powerful pulse shape analysis
techniques profiting from the p-type point contact HPGe detectors technology.
The effectiveness of these methods is assessed using simulations of the
different background components whose purity levels are constrained from
radioassay measurements. Preliminary background results obtained during the
engineering runs of the Demonstrator are presented.
-
We report the first measurement of the total MUON flux underground at the
Davis Campus of the Sanford Underground Research Facility at the 4850 ft level.
Measurements were done with the Majorana Demonstrator veto system arranged in
two different configurations. The measured total flux is (5.31+/-0.17) x 10^-9
muons/s/cm^2.
-
The MAJORANA Collaboration is constructing the MAJORANA DEMONSTRATOR, an
ultra-low background, modular, HPGe detector array with a mass of 44.8-kg (29.7
kg enriched >88% in Ge-76) to search for neutrinoless double beta decay in
Ge-76. The next generation of tonnescale Ge-based neutrinoless double beta
decay searches will probe the neutrino mass scale in the inverted-hierarchy
region. The MAJORANA DEMONSTRATOR is envisioned to demonstrate a path forward
to achieve a background rate at or below 1 count/tonne/year in the 4 keV region
of interest around the Q-value of 2039 keV. The MAJORANA DEMONSTRATOR follows a
modular implementation to be easily scalable to the next generation experiment.
First data taken with the DEMONSTRATOR are introduced here.
-
The MAJORANA collaboration is constructing the MAJORANA DEMONSTATOR at the
Sanford Underground Research Facility at the Homestake gold mine, in Lead, SD.
The apparatus will use Ge detectors, enriched in isotope \nuc{76}{Ge}, to
demonstrate the feasibility of a large-scale Ge detector experiment to search
for neutrinoless double beta decay. The long half-life of this postulated
process requires that the apparatus be extremely low in radioactive isotopes
whose decays may produce backgrounds to the search. The radioassay program
conducted by the collaboration to ensure that the materials comprising the
apparatus are sufficiently pure is described. The resulting measurements of the
radioactive-isotope contamination for a number of materials studied for use in
the detector are reported.
-
The MAJORANA Collaboration is constructing the MAJORANA Demonstrator, an
ultra-low background, 44-kg modular high-purity Ge (HPGe) detector array to
search for neutrinoless double-beta decay in Ge-76. The phenomenon of surface
micro-discharge induced by high-voltage has been studied in the context of the
MAJORANA Demonstrator. This effect can damage the front-end electronics or
mimic detector signals. To ensure the correct performance, every high-voltage
cable and feedthrough must be capable of supplying HPGe detector operating
voltages as high as 5 kV without exhibiting discharge. R&D measurements were
carried out to understand the testing system and determine the optimum design
configuration of the high-voltage path, including different improvements of the
cable layout and feedthrough flange model selection. Every cable and
feedthrough to be used at the MAJORANA Demonstrator was characterized and the
micro-discharge effects during the MAJORANA Demonstrator commissioning phase
were studied. A stable configuration has been achieved, and the cables and
connectors can supply HPGe detector operating voltages without exhibiting
discharge.
-
The MAJORANA Collaboration is constructing the MAJORANA DEMONSTRATOR, an
ultra-low background, modular, HPGe detector array with a mass of 44-kg (29 kg
76Ge and 15 kg natGe) to search for neutrinoless double beta decay in Ge-76.
The next generation of tonne-scale Ge-based neutrinoless double beta decay
searches will probe the neutrino mass scale in the inverted-hierarchy region.
The MAJORANA DEMONSTRATOR is envisioned to demonstrate a path forward to
achieve a background rate at or below 1 count/tonne/year in the 4 keV region of
interest around the Q-value of 2039 keV. The MAJORANA DEMONSTRATOR follows a
modular implementation to be easily scalable to the next generation experiment.
First, the prototype module was assembled; it has been continuously taking data
from July 2014 to June 2015. Second, Module 1 with more than half of the total
enriched detectors and some natural detectors has been assembled and it is
being commissioned. Finally, the assembly of Module 2, which will complete
MAJORANA DEMONSTRATOR, is already in progress.
-
The MAJORANA Collaboration will seek neutrinoless double beta decay (0nbb) in
76Ge using isotopically enriched p-type point contact (PPC) high purity
Germanium (HPGe) detectors. A tonne-scale array of HPGe detectors would require
background levels below 1 count/ROI-tonne-year in the 4 keV region of interest
(ROI) around the 2039 keV Q-value of the decay. In order to demonstrate the
feasibility of such an experiment, the MAJORANA DEMONSTRATOR, a 40 kg HPGe
detector array, is being constructed with a background goal of <3
counts/ROI-tonne-year, which is expected to scale down to <1
count/ROI-tonne-year for a tonne-scale experiment. The signal readout
electronics, which must be placed in close proximity to the detectors, present
a challenge toward reaching this background goal. This talk will discuss the
materials and design used to construct signal readout electronics with low
enough backgrounds for the MAJORANA DEMONSTRATOR.
-
The MAJORANA Collaboration is constructing the MAJORANA DEMONSTRATOR, an
ultra-low background, 40-kg modular HPGe detector array to search for
neutrinoless double beta decay in 76Ge. In view of the next generation of
tonne-scale Ge-based 0nbb-decay searches that will probe the neutrino mass
scale in the inverted-hierarchy region, a major goal of the MAJORANA
DEMONSTRATOR is to demonstrate a path forward to achieving a background rate at
or below 1 count/tonne/year in the 4 keV region of interest around the Q-value
at 2039 keV. The background rejection techniques to be applied to the data
include cuts based on data reduction, pulse shape analysis, event coincidences,
and time correlations. The Point Contact design of the DEMONSTRATOR 0s
germanium detectors allows for significant reduction of gamma background.
-
The MAJORANA Collaboration is constructing the MAJORANA Demonstrator, an
ultra-low background, 40-kg modular high purity Ge detector array to search for
neutrinoless double-beta decay in Ge. In view of the next generation of
tonne-scale Ge-based neutrinoless double-beta decay searches that will probe
the neutrino mass scale in the inverted-hierarchy region, a major goal of the
Demonstrator is to demonstrate a path forward to achieving a background rate at
or below 1 count/tonne/year in the 4 keV region of interest around the Q-value
at 2039 keV. The current status of the Demonstrator is discussed, as are plans
for its completion.
-
The MAJORANA DEMONSTRATOR is an array of natural and enriched high purity
germanium detectors that will search for the neutrinoless double-beta decay of
76-Ge and perform a search for weakly interacting massive particles (WIMPs)
with masses below 10 GeV. As part of the MAJORANA research and development
efforts, we have deployed a modified, low-background broad energy germanium
detector at the Kimballton Underground Research Facility. With its sub-keV
energy threshold, this detector is sensitive to potential non-Standard Model
physics, including interactions with WIMPs. We discuss the backgrounds present
in the WIMP region of interest and explore the impact of slow surface event
contamination when searching for a WIMP signal.
-
A low-background, high-purity germanium detector has been used to search for
evidence of low-energy, bremsstrahlung-generated solar axions. An upper bound
of $1.36\times 10^{-11}$ $(95% CL)$ is placed on the direct coupling of DFSZ
model axions to electrons. The prospects for the sensitivity of the Majorana
Demonstrator array of point-contact germanium detectors to solar axions are
discussed in the context of the model-independent annual modulation due to the
seasonal variation of the earth-sun distance.
-
The MAJORANA DEMONSTRATOR neutrinoless double beta-decay experiment is
currently under construction at the Sanford Underground Research Facility in
South Dakota, USA. An overview and status of the experiment are given.
-
The {\sc Majorana Demonstrator will search for the neutrinoless double-beta
decay of the isotope Ge-76 with a mixed array of enriched and natural germanium
detectors. The observation of this rare decay would indicate the neutrino is
its own antiparticle, demonstrate that lepton number is not conserved, and
provide information on the absolute mass scale of the neutrino. The {\sc
Demonstrator} is being assembled at the 4850-foot level of the Sanford
Underground Research Facility in Lead, South Dakota. The array will be situated
in a low-background environment and surrounded by passive and active shielding.
Here we describe the science goals of the {\sc Demonstrator} and the details of
its design.