• A new method of pulse shape discrimination (PSD) for BEGe detectors is developed to suppress Compton-continuum by digital pulse shape analysis (PSA), which helps reduce the Compton background level in gamma ray spectrometry. A decision parameter related to the rise time of a pulse shape was presented. The method was verified by experiments using 60Co and 137Cs sources. The result indicated that the 60Co Peak to Compton ratio and the Cs-Peak to Co-Compton ratio could be improved by more than two and three times, respectively.
  • An underwater in situ gamma ray spectrometer based on LaBr3 was developed and optimized to monitor marine radioactivity. The intrinsic background mainly from La138 and Ac227 of LaBr3 was well determined by low background measurement and pulse shape discrimination method. A method of self-calibration using three internal contaminant peaks was proposed to eliminate the peak shift during long term monitoring. With experiments under different temperatures, the method was proved to be helpful for maintaining long term stability. To monitor the marine radioactivity, the spectrometer efficiency was calculated via water tank experiment as well as Monte Carlo simulation.
  • A low background germanium gamma ray spectrometer, GeTHU, has been installed at China JinPing underground Laboratory. The integral background count rate between 40 and 2700 keV was 0.6 cpm, and the origin was studied by Monte Carlo simulation. Detection limits and efficiencies were calculated for selected gamma peaks. Boric acid and silica sand samples were measured and 137Cs contamination was found in boric acid. GeTHU will be mainly used to measure environmental samples and screen materials in dark matter experiments.