Silicon drift detectors (SDDs) revolutionized spectroscopy in fields as
diverse as geology and dentistry. For a subset of experiments at ultra-fast,
x-ray free-electron lasers (FELs), SDDs can make substantial contributions.
Often the unknown spectrum is interesting, carrying science data, or the
background measurement is useful to identify unexpected signals. Many
measurements involve only several discrete photon energies known a priori,
allowing single event decomposition of pile-up and spectroscopic photon
counting. We designed a pulse function and demonstrated that the signal
amplitude and rise time are obtained for each pulse by fitting, thus removing
the need for pulse shaping. By avoiding pulse shaping, rise times of tens of
nanoseconds resulted in reduced pulse pile-up and allowed decomposition of
remaining pulse pile-up at photon separation times down to hundreds of
nanoseconds while yielding time-of-arrival information with precision of 10
nanoseconds. Waveform fitting yields simultaneously high energy resolution and
high counting rates (2 orders of magnitude higher than current digital pulse
processors). We showed that pile-up spectrum fitting is relatively simple and
preferable to pile-up spectrum deconvolution. We developed a photon pile-up
statistical model for constant intensity sources, extended it to variable
intensity sources (typical for FELs) and used it to fit a complex pile-up
spectrum. We subsequently developed a Bayesian pile-up decomposition method
that allows decomposing pile-up of single events with up to 6 photons from 6
monochromatic lines with 99% accuracy. The usefulness of SDDs will continue
into the x-ray FEL era of science. Their successors, the ePixS hybrid pixel
detectors, already offer hundreds of pixels, each with similar performance to
an SDD, in a compact, robust and affordable package
-
The search for dark matter is of fundamental importance to our understanding
of the universe. Weakly-Interacting Slim Particles (WISPs) such as axions and
hidden sector photons (HSPs) are well motivated candidates for the dark matter.
Some of the most sensitive and mature experiments to detect WISPs rely on
microwave cavities, and the detection of weak photon signals. It is often
suggested to power combine multiple cavities, which creates a host of technical
concerns. We outline a scheme based on cross-correlation for effectively power
combining cavities and increasing the signal-to-noise ratio of a candidate WISP
signal.
-
The motivation for a cosmic muon veto (CMV) detector is to explore the
possibility of locating the proposed large Iron Calorimeter (ICAL) detector at
the India based Neutrino Observatory (INO) at a shallow depth. An initial
effort in that direction, through the assembly and testing of a ∼ 1 m
× 1 m × 0.3 m plastic scintillator based detector, is described.
The plan for making a CMV detector for a smaller prototype mini-ICAL is also
outlined.
-
Spring-antispring systems have been investigated as possible low-frequency
seismic isolation in high-precision optical experiments. These systems provide
the possibility to tune the fundamental resonance frequency to, in principle,
arbitrarily low values, and at the same time maintain a compact design of the
isolation system. It was argued though that thermal noise in spring-antispring
systems would not be as small as one may naively expect from lowering the
fundamental resonance frequency. In this paper, we present a detailed
calculation of the suspension thermal noise for a specific spring-antispring
system, namely the Roberts linkage. We find a concise expression of the
suspension thermal noise spectrum, which assumes a form very similar to the
well-known expression for a simple pendulum. It is found that while the Roberts
linkage can provide strong seismic isolation due to a very low fundamental
resonance frequency, its thermal noise is rather determined by the dimension of
the system. We argue that this is true for all horizontal mechanical isolation
systems with spring-antispring dynamics. This imposes strict requirements on
mechanical spring-antispring systems for the seismic isolation in potential
future low-frequency gravitational-wave detectors as we discuss for the four
main concepts: atom-interferometric, superconducting, torsion-bars, and
conventional laser interferometer.
-
Slow liquid scintillator Cherenkov detectors have been proposed as part of
several future neutrino experiments because they can provide both
directionality and energy measurements. This feature is expected to enhance the
sensitivities for MeV-scale neutrino physics, including solar physics, the
search for supernova relic neutrino, and the study of geo-sciences. In this
study, the characteristics of a slow liquid scintillator were investigated,
along with the light yields and decay time constants for various combinations
of linear alkylbenzene (LAB), 2,5-diphenyloxazole (PPO), and 1,4-bis
(2-methylstyryl)-benzene (bis-MSB). The results of our study indicated that LAB
with 0.07 g/L of PPO and 13 mg/L of bis-MSB was the best candidate for an
effective separation between Cherenkov and scintillation lights with a
reasonably high light yield.
-
The SeaQuest spectrometer at Fermilab was designed to detect
oppositely-charged pairs of muons (dimuons) produced by interactions between a
120 GeV proton beam and liquid hydrogen, liquid deuterium and solid nuclear
targets. The primary physics program uses the Drell-Yan process to probe
antiquark distributions in the target nucleon. The spectrometer consists of a
target system, two dipole magnets and four detector stations. The upstream
magnet is a closed-aperture solid iron magnet which also serves as the beam
dump, while the second magnet is an open aperture magnet. Each of the detector
stations consists of scintillator hodoscopes and a high-resolution tracking
device. The FPGA-based trigger compares the hodoscope signals to a set of
pre-programmed roads to determine if the event contains oppositely-signed,
high-mass muon pairs.
-
The Hamamatsu R5912-HQE photomultiplier-tube (PMT) is a novel high-quantum
efficiency PMT. It is currently used in the DEAP-3600 dark matter detector and
is of significant interest for future dark matter and neutrino experiments
where high signal yields are needed.
We report on the methods developed for in-situ characterization and
monitoring of DEAP's 255 R5912-HQE PMTs. This includes a detailed discussion of
typical measured single-photoelectron charge distributions, correlated noise
(afterpulsing), dark noise, double, and late pulsing characteristics. The
characterization is performed during the detector commissioning phase using
laser light injected through a light diffusing sphere and during normal
detector operation using LED light injected through optical fibres.
-
Modern TOF-PET scanner systems require high-speed computing resources for
efficient data processing, monitoring and image reconstruction. In this article
we present the data flow and software architecture for the novel TOF-PET
scanner developed by the J-PET collaboration. We discuss the data acquisition
system, reconstruction framework and image reconstruction software. Also, the
concept of computing outside hospitals in the remote centers such as \'Swierk
Computing Centre in Poland is presented.
-
The Transition Radiation Detector (TRD) was designed and built to enhance the
capabilities of the ALICE detector at the Large Hadron Collider (LHC). While
aimed at providing electron identification and triggering, the TRD also
contributes significantly to the track reconstruction and calibration in the
central barrel of ALICE. In this paper the design, construction, operation, and
performance of this detector are discussed. A pion rejection factor of up to
410 is achieved at a momentum of 1 GeV/c in p-Pb collisions and the
resolution at high transverse momentum improves by about 40% when including the
TRD information in track reconstruction. The triggering capability is
demonstrated both for jet, light nuclei, and electron selection.
-
We examine the sensitivity of a large scale two-phase liquid argon detector
to the directionality of the dark matter signal. This study was performed under
the assumption that, above 50 keV of recoil energy, one can determine (with
some resolution) the direction of the recoil nucleus without head-tail
discrimination, as suggested by past studies that proposed to exploit the
dependence of columnar recombination on the angle between the recoil nucleus
direction and the electric field. In this paper we study the differential
interaction recoil rate as a function of the recoil direction angle with
respect to the zenith for a detector located at the Laboratori Nazionali del
Gran Sasso and we determine its diurnal and seasonal modulation. Using a
likelihood-ratio based approach we show that, with the angular information
alone, 100 events are enough to reject the isotropic hypothesis at three
standard deviation level. For an exposure of 100 tonne years this would
correspond to a spin independent WIMP-nucleon cross section of about 10^-46
cm^2 at 200 GeV WIMP mass. The results presented in this paper provide strong
motivation for the experimental determination of directional recoil effects in
two-phase liquid argon detectors.
-
Far-infrared astronomy has advanced rapidly since its inception in the late
1950's, driven by a maturing technology base and an expanding community of
researchers. This advancement has shown that observations at far-infrared
wavelengths are important in nearly all areas of astrophysics, from the search
for habitable planets and the origin of life, to the earliest stages of galaxy
assembly in the first few hundred million years of cosmic history. The
combination of a still developing portfolio of technologies, particularly in
the field of detectors, and a widening ensemble of platforms within which these
technologies can be deployed, means that far-infrared astronomy holds the
potential for paradigm-shifting advances over the next decade. In this review,
we examine current and future far-infrared observing platforms, including
ground-based, sub-orbital, and space-based facilities, and discuss the
technology development pathways that will enable and enhance these platforms to
best address the challenges facing far-infrared astronomy in the 21st century.
-
Within the framework of xenon-based double beta decay experiments, we propose
the possibility to improve the background rejection of an electroluminescent
Time Projection Chamber (EL TPC) by reducing the diffusion of the drifting
electrons while keeping nearly intact the energy resolution of a pure xenon EL
TPC. Based on state-of-the-art microscopic simulations, a substantial addition
of helium, around 10 or 15~\%, may reduce drastically the transverse diffusion
down to 2.5~mm/√m from the 10.5~mm/√m of pure
xenon. The longitudinal diffusion remains around 4~mm/√m.
Light production studies have been performed as well. They show that the
relative variation in energy resolution introduced by such a change does not
exceed a few percent, which leaves the energy resolution practically unchanged.
The technical caveats of using photomultipliers close to an helium atmosphere
are also discussed in detail.
-
This paper presents the standardization of Proton Induced X-rays Emission
(PIXE) technique for the trace element analysis of thick standard samples.
Three standard reference materials (SRMs) viz-\`a-vis titanium, copper and iron
base alloys were used for the study due to their availability. The protons beam
was accelerated up to 2.57 MeV energy by 5UDH-II tandem Pelletron accelerator
and samples were irradiated at different geometry and durations. Spectrum was
acquired using a multi-channel spectrum analyzer while spectrum analysis was
done using a GUPIXWIN model for determination of elemental concentrations of
trace elements. The obtained experimental data was compared with theoretical
data and results were found in close agreement.
-
The super Pioneering High Energy Nuclear Interaction eXperiment (sPHENIX) at
the Relativistic Heavy Ion Collider (RHIC) will perform high precision
measurements of jets and heavy flavor observables for a wide selection of
nuclear collision systems, elucidating the microscopic nature of strongly
interacting matter ranging from nucleons to the strongly coupled quark-gluon
plasma. A prototype of the sPHENIX calorimeter system was tested at the
Fermilab Test Beam Facility as experiment T-1044 in the spring of 2016. The
electromagnetic calorimeter (EMCal) prototype is composed of scintillating
fibers embedded in a mixture of tungsten powder and epoxy. The hadronic
calorimeter (HCal) prototype is composed of tilted steel plates alternating
with plastic scintillator. Results of the test beam reveal the energy
resolution for electrons in the EMCal is 2.8%⊕ 15.5%/√E and the
energy resolution for hadrons in the combined EMCal plus HCal system is
13.5%⊕64.9%/√E. These results demonstrate that the performance
of the proposed calorimeter system satisfies the sPHENIX specifications.
-
We present a particle-level model for calculating the radio scatter of
incident RF radiation from the plasma formed in the wake of a particle shower.
We incorporate this model into a software module ("RadioScatter"), which
calculates the collective scattered signal using the individual particle
equations of motion, accounting for plasma effects, transmitter and receiver
geometries, refraction at boundaries, and antenna gain patterns. We find
appreciable collective scattering amplitudes with coherent phase for a range of
geometries, with high geometric and volumetric acceptance. Details of the
calculation are discussed, as well as the implementation of RadioScatter into
GEANT4. A laboratory test of our model, currently scheduled at SLAC in 2018,
with the goal of measuring the time-dependent characteristics of the reflecting
plasma, is also described. Prospects for a future in-ice, high-energy neutrino
detector, along with comparison to current detection strategies, are presented.
-
The authors' design for a M\"ossbauer spectroscopy unit accessed via the
Internet is presented. The spectrometer's configuration, operational
peculiarities, key specifications, and remote access procedures are described.
More detailed educational, scientific, and technical information can be found
at the project's official website http://efmsb.sinp.msu.ru.
-
Reentrant cavities are microwave resonant devices employed in a number of
different areas of physics. They are appealing due to their simple frequency
tuning mechanism, which offers large tuning ranges. Reentrant cavities are, in
essence, 3D lumped LC circuits consisting of a conducting central post embedded
in a resonant cavity. The lowest order reentrant mode (which transforms from
the TM010 mode) has been extensively studied in past publications. In this
work we show the existence of higher order reentrant post modes (which
transform from the TM01n mode family). We characterize these new modes in
terms of their frequency tuning, filling factors and quality factors, as well
as discuss some possible applications of these modes in fundamental physics
tests. The appendix contains a comment on a paper related to this work.
-
We present a concept of an electromagnetic phase-shift flowmeter that has a
significantly reduced sensitivity to the variation of the electrical
conductivity of a liquid metal. A simple theoretical model of the flowmeter is
considered where the flow is approximated by a solid finite-thickness
conducting layer which moves in the presence of an ac magnetic field. In
contrast to the original design [Priede et al., Meas. Sci. Technol. 22 (2011)
055402], where the flow rate is determined by measuring only the phase shift
between the voltages induced in two receiving coils, the improved design
measures also the phase shift between the sending and the upstream receiving
coils. These two phase shifts are referred to as internal and external ones,
respectively. We show that the effect of electrical conductivity on the
internal phase shift, which is induced by the flow, can be strongly reduced by
rescaling it with the external phase shift, which depends mostly on the
conductivity of medium. Two different rescalings are found depending on the ac
frequency. At low frequencies, when the shielding effect is weak, the effect of
conductivity is strongly reduced by rescaling the internal phase shift with the
external one squared. At higher frequencies, the same is achieved by rescaling
the internal phase shift directly with the external one.
-
This work presents a study on the fundamental aspects of a reliable gold
determination in voluminous mineral samples by Neutron Activation Analysis,
using a Linear Particle Accelerator as a source of irradiation and a single
detector as a detection system. The report includes i) an evaluation of the
detection efficiency of the system, particularly the evaluation of the
geometric efficiency in samples with significant volumes (with approximate
volumes from 20 to 30 cm3); ii) a few methods for the calculation of the gamma
ray self attenuation in the sample; iii) some gold in-homogeneity tests in the
sample; iv) several validation calculations obtained with the MCNP code - the
complete system was modeled with this program. This group of results has
produced a gamma-spectrometry protocol for this determination, in order to
obtain accurately the gold concentration in mineral samples this study was
required by a legal bureau to determine the gold abundance in a Patagonian
small deposit. According to the high quality of the results obtained from the
process of characterization of this method, it is concluded that this gold
analysis would be obtained with an accuracy similar than the one provided by
the usual NAA using a reactor, in small samples exactitude values around few %.
-
We explore the application of heterodyne interferometry for a weak-field
coherent detection scheme. The methods detailed here will be used in ALPS II,
an experiment designed to search for weakly-interacting, sub-eV particles. For
ALPS II to reach its design sensitivity this detection system must be capable
of accurately measuring fields with equivalent amplitudes on the order of
10−5 photons per second or greater. We present initial results of an
equivalent dark count rate on the order of 10−5 photons per second as well
as successful generation and detection of a signal with a field strength
equivalent to 10−2 photons per second.
-
The evolution of particle detectors dates back to the discovery of X-rays and
radioactivity in 1890s. In detector history, the Resistive Plate Chambers
(RPCs) are introduced in early 1980s. An RPC is a gaseous detector made up of
two parallel electrodes having high resistivity like that of glass and
bakelite. Currently several high energy physics experiments are using RPC-based
detector system due to robustness and simplicity of construction. In each and
every experiment, RPCs have to run continuously for several years. So, it
demands an in-depth characterization of the electrode materials. In the present
study, an elemental analysis of locally available glass and bakelite samples is
done using PIXE facility available at Panjab University Cyclotron, Chandigarh.
PIXE measurements are done using 2.7 MeV proton beam incident on the electrode
sample target. The constituent elements present in these electrode samples are
reported.
-
We present a comprehensive numerical analysis of alternative transient
eddy-current flowmetering methods for liquid metals. This type of flowmeter
operates by tracking eddy-current markers excited by the magnetic field pulses
in the flow of a conducting liquid. Using a simple mathematical model, where
the fluid flow is replaced by a translating cylinder, a number possible
alternative measurement schemes are considered. The velocity of the medium can
be measured by tracking zero crossing points and spatial or temporal extrema of
the electromotive force (emf) induced by transient eddy currents in the
surrounding space. Zero crossing points and spatial extrema of the emf travel
synchronously with the medium whereas temporal extrema experience an initial
time delay which depends on the conductivity and velocity of the medium.
Performance of transient eddy-current flowmetering depends crucially on the
symmetry of system. Eddy current asymmetry of a few per cent makes the
detection point drift with a velocity corresponding to a magnetic Reynolds
number Rm∼0.1. With this level of asymmetry transient eddy-current
flowmetering can be reliably applicable only to flows with Rm≳0.1. A
more accurate symmetry adjustment or calibration of flowmeters may be necessary
at lower velocities.
-
Light-shining-through-a-wall experiments represent a new experimental
approach in the search for undiscovered elementary particles not accessible
with accelerator based experiments. The next generation of these experiments,
such as ALPS II, require high finesse, long baseline optical cavities with fast
length control. In this paper we report on a length stabilization control loop
used to keep a 9.2 m cavity resonant. The finesse of this cavity was measured
to be 101,300±500 for 1064 nm light. Fluctuations in the differential
cavity length as seen with 1064 nm and 532 nm light were measured. Such
fluctuations are of high relevance, since 532 nm light will be used to sense
the length of the ALPS II regeneration cavity. Limiting noise sources and
different control strategies are discussed, in order to fulfill the length
stability requirements for ALPS II.
-
We report on recent advances in the operation of bubble-assisted Liquid Hole
Multipliers (LHM). By confining a vapor bubble under or adjacent to a
perforated electrode immersed in liquid xenon, we could record both
radiation-induced ionization electrons and primary scintillation photons in the
noble liquid. Four types of LHM electrodes were investigated: a THGEM, standard
double-conical GEM, 50 μm-thick single-conical GEM (SC-GEM) and 125
μm-thick SC-GEM - all coated with CsI photocathodes. The 125 μm-thick
SC-GEM provided the highest electroluminescence (EL) yields, up to ~400 photons
per electron over 4π with an RMS pulse-height resolution reaching 5.5% for
events comprising ~7000 primary electrons. Applying a high transfer field
across the bubble, the EL yield was further increased by a factor of ~5. The
feasibility of a vertical-mode LHM, with the bubble confined between two
vertical electrodes, and the operation of a two-stage LHM configuration were
demonstrated for the first time. We combine electrostatic simulations with
observed signals to draw conclusions regarding the location of the liquid-gas
interface and suggest an explanation for the observed differences in EL yield
between the investigated electrodes.
-
Free-space optical communications have held the promise of revolutionizing
space communications for a long time. The benefits of increasing the bitrate
while reducing the volume, mass and energy of the space terminals have
attracted the attention of many researchers for a long time. In the last few
years, more and more technology demonstrations have been taking place with
participants from both the public and the private sector. The National
Institute of Information and Communications Technology (NICT) in Japan has a
long experience in this field. SOTA (Small Optical TrAnsponder) was the last
NICT space lasercom mission, designed to demonstrate the potential of this
technology applied to microsatellites. Since the beginning of SOTA mission in
2014, NICT regularly established communication using the Optical Ground
Stations (OGS) located in the Headquarters at Koganei (Tokyo) to receive the
SOTA signals, with over one hundred successful links. All the goals of the SOTA
mission were fulfilled, including up to 10-Mbit/s downlinks using two different
wavelengths and apertures, coarse and fine tracking of the OGS beacon,
space-to-ground transmission of the on-board-camera images, experiments with
different error correcting codes, interoperability with other international
OGS, and experiments on quantum communications. The SOTA mission ended on
November 2016, more than doubling the designed lifetime of 1-year. In this
paper, the SOTA characteristics and basic operation are explained, along with
the most relevant technological demonstrations.