• A ThO$_{2}$ sample and a nickel activation foil were irradiated in the leakage neutron field of CFBR-II reactor. The activities of the activation products were measured after irradiation to obtain the reaction rates. The normalized reaction rates were also calculated based on the ENDF/B-VII.1, CENDL-3.1, JENDL-4.0, BROND-2.2 databases. The experimental reaction rate ratio is 4.37 with an uncertainty of 3.9\% which is coincident with each of the ratios calculated based on the ENDFB-VII. 1, JENDL-4.0, BROND-2.2 databases, but is 11.2\% larger than that based on CENDL-3.1 database.
  • The Pu-239 fission chambers are widely used to measure fission spectrum neutron flux due to a flat response to fast neutrons. However, in the mean time the resonance and thermal neutrons can cause a significant influence on the measurement if they are moderated, which could be eliminated by using B and Cd covers. At a column enriched uranium fast neutron critical assembly, the fission reaction rates of Pu-239 are measured as 1.791*10-16,2.350*10-16 and 1.385*10-15 per second for 15mm thick B cover, 0.5mm thick Cd cover, and no cover respectively. While the fission reaction rate of Pu-239 is rapidly increased to 2.569*10-14 for a 20mm thick polythene covering fission chamber. The average Pu-239 fission cross-section of thermal and resonance neutrons is calculated to be 500b and 24.95b with the assumption of 1/v and 1/E spectra respectively, then thermal, resonance and fast neutron flux are achieved to be 2.30*106,2.24*106 and 1.04*108cm-2s-1.