We present the precision calibration of 35 Hamamatsu R11410-22
photomultiplier tubes (PMTs) with xenon scintillation light centred near 175
nm. This particular PMT variant was developed specifically for the LUX-ZEPLIN
(LZ) dark matter experiment. A room-temperature xenon scintillation cell
coupled to a vacuum cryostat was used to study the full-face PMT response at
both room and low temperature $\textrm{($\sim$ -100$^\circ$C)}$, in particular
to determine the quantum efficiency (QE) and double photoelectron emission
(DPE) probability in LZ operating conditions. For our sample with an average QE
of $\textrm{(32.4$\pm$2.9)%}$ at room temperature, we find a relative
improvement of $\textrm{(17.9$\pm$5.2)%}$ upon cooling (where uncertainty
values refer to the sample standard deviation). The mean DPE probability in
response to single vacuum ultraviolet (VUV) photons is
$\textrm{(22.6$\pm$2.0)%}$ at low temperature; the DPE increase relative to
room temperature, measured here for the first time, was
$\textrm{(12.2$\pm$3.9)%}$. Evidence of a small triple photoelectron emission
probability $\textrm{($\sim$0.6%)}$ has also been observed. Useful correlations
are established between these parameters and the QE as measured by the
manufacturer. The single VUV photon response is also measured for one ETEL
D730/9829QB, a PMT with a more standard bialkali photocathode used in the
ZEPLIN-III experiment, for which we obtained a cold DPE fraction of
$\textrm{(9.1$\pm$0.1)%}$. Hence, we confirm that this effect is not restricted
to the low-temperature bialkali photocathode technology employed by Hamamatsu.
This highlights the importance of considering this phenomenon in the
interpretation of data from liquid xenon scintillation and electroluminescence
detectors, and from many other optical measurements in this wavelength region.
-
The $(x, y)$ position reconstruction method used in the analysis of the
complete exposure of the Large Underground Xenon (LUX) experiment is presented.
The algorithm is based on a statistical test that makes use of an iterative
method to recover the photomultiplier tube (PMT) light response directly from
the calibration data. The light response functions make use of a two
dimensional functional form to account for the photons reflected on the inner
walls of the detector. To increase the resolution for small pulses, a photon
counting technique was employed to describe the response of the PMTs. The
reconstruction was assessed with calibration data including
${}^{\mathrm{83m}}$Kr (releasing a total energy of 41.5 keV) and ${}^{3}$H
($\beta^-$ with Q = 18.6 keV) decays, and a deuterium-deuterium (D-D) neutron
beam (2.45 MeV). In the horizontal plane, the reconstruction has achieved an
$(x, y)$ position uncertainty of $\sigma$= 0.82 cm for events of only 200
electroluminescence photons and $\sigma$ = 0.17 cm for 4,000
electroluminescence photons. Such signals are associated with electron recoils
of energies $\sim$0.25 keV and $\sim$10 keV, respectively. The reconstructed
position of the smallest events with a single electron emitted from the liquid
surface has a horizontal $(x, y)$ uncertainty of 2.13 cm.
-
Noble liquid radiation detectors have long been afflicted by spurious
electron emission from their cathodic electrodes. This phenomenon must be
understood and mitigated in the next generation of liquid xenon (LXe)
experiments searching for WIMP dark matter or neutrinoless double beta decay,
and in the large liquid argon (LAr) detectors for the long-baseline neutrino
programme. We present a systematic study of this spurious emission involving a
series of slow voltage-ramping tests on fine metal wires immersed in a
two-phase xenon time projection chamber with single electron sensitivity.
Emission currents as low as $10^{-18}$ A can thus be detected by electron
counting, a vast improvement over previous dedicated measurements. Emission
episodes were recorded at surface fields as low as $\sim$10 kV/cm in some wires
and observed to have complex emission patterns, with average rates of 10-200
c/s and outbreaks as high as $\sim$10$^6$ c/s. A fainter, less variable type of
emission was also present in all untreated samples. There is evidence of a
partial conditioning effect, with subsequent tests yielding on average fewer
emitters occurring at different fields for the same wire. We find no evidence
for an intrinsic threshold particular to the metal-LXe interface which might
have limited previous experiments up to fields of at least 160 kV/cm. The
general phenomenology is not consistent with enhanced field emission from
microscopic filaments, but it appears instead to be related to the quality of
the wire surface in terms of corrosion and the nature of its oxide layer. This
study concludes that some surface treatments, in particular nitric acid
cleaning for stainless steel wires, can bringing about at least
order-of-magnitude improvements in overall electron emission rates, and this
should help the next generation of detectors achieve the required electrostatic
performance.
-
This work details the development of a three-dimensional (3D) electric field
model for the LUX detector. The detector took data during two periods of
searching for weakly interacting massive particle (WIMP) searches. After the
first period completed, a time-varying non-uniform negative charge developed in
the polytetrafluoroethylene (PTFE) panels that define the radial boundary of
the detector's active volume. This caused electric field variations in the
detector in time, depth and azimuth, generating an electrostatic
radially-inward force on electrons on their way upward to the liquid surface.
To map this behavior, 3D electric field maps of the detector's active volume
were built on a monthly basis. This was done by fitting a model built in COMSOL
Multiphysics to the uniformly distributed calibration data that were collected
on a regular basis. The modeled average PTFE charge density increased over the
course of the exposure from -3.6 to $-5.5~\mu$C/m$^2$. From our studies, we
deduce that the electric field magnitude varied while the mean value of the
field of $\sim200$~V/cm remained constant throughout the exposure. The varying
electric fields and their impact on event reconstruction and discrimination
were successfully modeled.
-
We report an absolute calibration of the ionization yields($\textit{Q$_y$})$
and fluctuations for electronic recoil events in liquid xenon at discrete
energies between 186 eV and 33.2 keV. The average electric field applied across
the liquid xenon target is 180 V/cm. The data are obtained using low energy
$^{127}$Xe electron capture decay events from the 95.0-day first run from LUX
(WS2013) in search of Weakly Interacting Massive Particles (WIMPs). The
sequence of gamma-ray and X-ray cascades associated with $^{127}$I
de-excitations produces clearly identified 2-vertex events in the LUX detector.
We observe the K- (binding energy, 33.2 keV), L- (5.2 keV), M- (1.1 keV), and
N- (186 eV) shell cascade events and verify that the relative ratio of observed
events for each shell agrees with calculations. The N-shell cascade analysis
includes single extracted electron (SE) events and represents the lowest-energy
electronic recoil $\textit{in situ}$ measurements that have been explored in
liquid xenon.
-
LUX was the first dark matter experiment to use a $^{83\textrm{m}}$Kr
calibration source. In this paper we describe the source preparation and
injection. We also present several $^{83\textrm{m}}$Kr calibration applications
in the context of the 2013 LUX exposure, including the measurement of temporal
and spatial variation in scintillation and charge signal amplitudes, and
several methods to understand the electric field within the time projection
chamber.
-
We present experimental constraints on the spin-dependent WIMP-nucleon
elastic cross sections from the total 129.5 kg-year exposure acquired by the
Large Underground Xenon experiment (LUX), operating at the Sanford Underground
Research Facility in Lead, South Dakota (USA). A profile likelihood ratio
analysis allows 90% CL upper limits to be set on the WIMP-neutron (WIMP-proton)
cross section of $\sigma_n$ = 1.6$\times 10^{-41}$ cm$^{2}$ ($\sigma_p$ =
5$\times 10^{-40}$ cm$^{2}$) at 35 GeV$c^{-2}$, almost a sixfold improvement
over the previous LUX spin-dependent results. The spin-dependent WIMP-neutron
limit is the most sensitive constraint to date.
-
We report constraints on spin-independent weakly interacting massive particle
(WIMP)-nucleon scattering using a 3.35e4 kg-day exposure of the Large
Underground Xenon (LUX) experiment. A dual-phase xenon time projection chamber
with 250 kg of active mass is operated at the Sanford Underground Research
Facility under Lead, South Dakota (USA). With roughly fourfold improvement in
sensitivity for high WIMP masses relative to our previous results, this search
yields no evidence of WIMP nuclear recoils. At a WIMP mass of 50 GeV/c^2,
WIMP-nucleon spin-independent cross sections above 2.2e-46 cm^2 are excluded at
the 90% confidence level. When combined with the previously reported LUX
exposure, this exclusion strengthens to 1.1e-46 cm^2 at 50 GeV/c^2.
-
The Large Underground Xenon (LUX) experiment is a dual-phase liquid xenon
time projection chamber (TPC) operating at the Sanford Underground Research
Facility in Lead, South Dakota. A calibration of nuclear recoils in liquid
xenon was performed $\textit{in situ}$ in the LUX detector using a collimated
beam of mono-energetic 2.45 MeV neutrons produced by a deuterium-deuterium
(D-D) fusion source. The nuclear recoil energy from the first neutron scatter
in the TPC was reconstructed using the measured scattering angle defined by
double-scatter neutron events within the active xenon volume. We measured the
absolute charge ($Q_{y}$) and light ($L_{y}$) yields at an average electric
field of 180 V/cm for nuclear recoil energies spanning 0.7 to 74 keV and 1.1 to
74 keV, respectively. This calibration of the nuclear recoil signal yields will
permit the further refinement of liquid xenon nuclear recoil signal models and,
importantly for dark matter searches, clearly demonstrates measured ionization
and scintillation signals in this medium at recoil energies down to
$\mathcal{O}$(1 keV).
-
We present the first experimental constraints on the spin-dependent
WIMP-nucleon elastic cross sections from LUX data acquired in 2013. LUX is a
dual-phase xenon time projection chamber operating at the Sanford Underground
Research Facility (Lead, South Dakota), which is designed to observe the recoil
signature of galactic WIMPs scattering from xenon nuclei. A profile likelihood
ratio analysis of $1.4~\times~10^{4}~\text{kg}\cdot~\text{days}$ of fiducial
exposure allows 90% CL upper limits to be set on the WIMP-neutron (WIMP-proton)
cross section of $\sigma_n~=~9.4~\times~10^{-41}~\text{cm}^2$
($\sigma_p~=~2.9~\times~10^{-39}~\text{cm}^2$) at 33 GeV/c$^2$. The
spin-dependent WIMP-neutron limit is the most sensitive constraint to date.
-
We present constraints on weakly interacting massive particles (WIMP)-nucleus
scattering from the 2013 data of the Large Underground Xenon dark matter
experiment, including $1.4\times10^{4}\;\mathrm{kg\; day}$ of search exposure.
This new analysis incorporates several advances: single-photon calibration at
the scintillation wavelength, improved event-reconstruction algorithms, a
revised background model including events originating on the detector walls in
an enlarged fiducial volume, and new calibrations from decays of an injected
tritium $\beta$ source and from kinematically constrained nuclear recoils down
to 1.1 keV. Sensitivity, especially to low-mass WIMPs, is enhanced compared to
our previous results which modeled the signal only above a 3 keV minimum
energy. Under standard dark matter halo assumptions and in the mass range above
4 $\mathrm{GeV}\,c^{-2}$, these new results give the most stringent direct
limits on the spin-independent WIMP-nucleon cross section. The 90% C.L. upper
limit has a minimum of 0.6 zb at 33 $\mathrm{GeV}\,c^{-2}$ WIMP mass.
-
We present measurements of the electron-recoil (ER) response of the LUX dark
matter detector based upon 170,000 highly pure and spatially-uniform tritium
decays. We reconstruct the tritium energy spectrum using the combined energy
model and find good agreement with expectations. We report the average charge
and light yields of ER events in liquid xenon at 180 V/cm and 105 V/cm and
compare the results to the NEST model. We also measure the mean charge
recombination fraction and its fluctuations, and we investigate the location
and width of the LUX ER band. These results provide input to a re-analysis of
the LUX Run3 WIMP search.
-
The design and performance of the LUX-ZEPLIN (LZ) detector is described as of
March 2015 in this Conceptual Design Report. LZ is a second-generation
dark-matter detector with the potential for unprecedented sensitivity to weakly
interacting massive particles (WIMPs) of masses from a few GeV/c2 to hundreds
of TeV/c2. With total liquid xenon mass of about 10 tonnes, LZ will be the most
sensitive experiment for WIMPs in this mass region by the end of the decade.
This report describes in detail the design of the LZ technical systems.
Expected backgrounds are quantified and the performance of the experiment is
presented. The LZ detector will be located at the Sanford Underground Research
Facility in South Dakota. The organization of the LZ Project and a summary of
the expected cost and current schedule are given.
-
We present the design and preliminary calibration results of a novel highly
miniaturised particle radiation monitor (HMRM) for spacecraft use. The HMRM
device comprises a telescopic configuration of active pixel sensors enclosed in
a titanium shield, with an estimated total mass of 52 g and volume of 15
cm$^3$. The monitor is intended to provide real-time dosimetry and
identification of energetic charged particles in fluxes of up to 10$^8$
cm$^{-2}$ s$^{-1}$ (omnidirectional). Achieving this capability with such a
small instrument could open new prospects for radiation detection in space.
-
We studied the application of statistical reconstruction algorithms, namely
maximum likelihood and least squares methods, to the problem of event
reconstruction in a dual phase liquid xenon detector. An iterative method was
developed for in-situ reconstruction of the PMT light response functions from
calibration data taken with an uncollimated gamma-ray source. Using the
techniques described, the performance of the ZEPLIN-III dark matter detector
was studied for 122 keV gamma-rays. For the inner part of the detector (R<100
mm), spatial resolutions of 13 mm and 1.6 mm FWHM were measured in the
horizontal plane for primary and secondary scintillation, respectively. An
energy resolution of 8.1% FWHM was achieved at that energy. The possibility of
using this technique for improving performance and reducing cost of
scintillation cameras for medical applications is currently under study.
-
We report experimental upper limits on WIMP-nucleon elastic scattering cross
sections from the second science run of ZEPLIN-III at the Boulby Underground
Laboratory. A raw fiducial exposure of 1,344 kg.days was accrued over 319 days
of continuous operation between June 2010 and May 2011. A total of eight events
was observed in the signal acceptance region in the nuclear recoil energy range
7-29 keV, which is compatible with background expectations. This allows the
exclusion of the scalar cross-section above 4.8E-8 pb near 50 GeV/c^2 WIMP mass
with 90% confidence. Combined with data from the first run, this result
improves to 3.9E-8 pb. The corresponding WIMP-neutron spin-dependent
cross-section limit is 8.0E-3 pb. The ZEPLIN programme reaches thus its
conclusion at Boulby, having deployed and exploited successfully three liquid
xenon experiments of increasing reach.
-
ZEPLIN-III is a two-phase xenon direct dark matter experiment located at the
Boulby Mine (UK). After its first science run in 2008 it was upgraded with: an
array of low background photomultipliers, a new anti-coincidence detector
system with plastic scintillator and an improved calibration system. After 319
days of data taking the second science run ended in May 2011. In this paper we
describe the instrument performance with emphasis on the position and energy
reconstruction algorithm and summarise the final science results.
-
Plastic scintillators are widely used in industry, medicine and scientific
research, including nuclear and particle physics. Although one of their most
common applications is in neutron detection, experimental data on their
response to low-energy nuclear recoils are scarce. Here, the relative
scintillation efficiency for neutron-induced nuclear recoils in a
polystyrene-based plastic scintillator (UPS-923A) is presented, exploring
recoil energies between 125 keV and 850 keV. Monte Carlo simulations,
incorporating light collection efficiency and energy resolution effects, are
used to generate neutron scattering spectra which are matched to observed
distributions of scintillation signals to parameterise the energy-dependent
quenching factor. At energies above 300 keV the dependence is reasonably
described using the semi-empirical formulation of Birks and a kB factor of
(0.014+/-0.002) g/MeVcm^2 has been determined. Below that energy the measured
quenching factor falls more steeply than predicted by the Birks formalism.
-
Scintillation and ionisation yields for nuclear recoils in liquid xenon above
10 keVnr (nuclear recoil energy) are deduced from data acquired using broadband
Am-Be neutron sources. The nuclear recoil data from several exposures to two
sources were compared to detailed simulations. Energy-dependent scintillation
and ionisation yields giving acceptable fits to the data were derived.
Efficiency and resolution effects are treated using a light collection Monte
Carlo, measured photomultiplier response profiles and hardware trigger studies.
A gradual fall in scintillation yield below ~40 keVnr is found, together with a
rising ionisation yield; both are in good agreement with the latest independent
measurements. The analysis method is applied to both the most recent ZEPLIN-III
data, acquired with a significantly upgraded detector and a
precision-calibrated Am-Be source, as well as to the earlier data from the
first run in 2008. A new method for deriving the recoil scintillation yield,
which includes sub-threshold S1 events, is also presented which confirms the
main analysis.
-
We present an experimental study of single electron emission in ZEPLIN-III, a
two-phase xenon experiment built to search for dark matter WIMPs, and discuss
applications enabled by the excellent signal-to-noise ratio achieved in
detecting this signature. Firstly, we demonstrate a practical method for
precise measurement of the free electron lifetime in liquid xenon during normal
operation of these detectors. Then, using a realistic detector response model
and backgrounds, we assess the feasibility of deploying such an instrument for
measuring coherent neutrino-nucleus elastic scattering using the ionisation
channel in the few-electron regime. We conclude that it should be possible to
measure this elusive neutrino signature above an ionisation threshold of
$\sim$3 electrons both at a stopped pion source and at a nuclear reactor.
Detectable signal rates are larger in the reactor case, but the triggered
measurement and harder recoil energy spectrum afforded by the accelerator
source enable lower overall background and fiducialisation of the active
volume.
-
We examine electron and nuclear recoil backgrounds from radioactivity in the
ZEPLIN-III dark matter experiment at Boulby. The rate of low-energy electron
recoils in the liquid xenon WIMP target is 0.75$\pm$0.05 events/kg/day/keV,
which represents a 20-fold improvement over the rate observed during the first
science run. Energy and spatial distributions agree with those predicted by
component-level Monte Carlo simulations propagating the effects of the
radiological contamination measured for materials employed in the experiment.
Neutron elastic scattering is predicted to yield 3.05$\pm$0.5 nuclear recoils
with energy 5-50 keV per year, which translates to an expectation of 0.4 events
in a 1-year dataset in anti-coincidence with the veto detector for realistic
signal acceptance. Less obvious background sources are discussed, especially in
the context of future experiments. These include contamination of scintillation
pulses with Cherenkov light from Compton electrons and from $\beta$ activity
internal to photomultipliers, which can increase the size and lower the
apparent time constant of the scintillation response. Another challenge is
posed by multiple-scatter $\gamma$-rays with one or more vertices in regions
that yield no ionisation. If the discrimination power achieved in the first run
can be replicated, ZEPLIN-III should reach a sensitivity of $\sim 1 \times
10^{-8}$ pb$\cdot$year to the scalar WIMP-nucleon elastic cross-section, as
originally conceived.
-
The design, optimisation and construction of an anti-coincidence veto
detector to complement the ZEPLIN-III direct dark matter search instrument is
described. One tonne of plastic scintillator is arranged into 52 bars
individually read out by photomultipliers and coupled to a gadolinium-loaded
passive polypropylene shield. Particular attention has been paid to
radiological content. The overall aim has been to achieve a veto detector of
low threshold and high efficiency without the creation of additional background
in ZEPLIN-III, all at a reasonable cost. Extensive experimental measurements of
the components have been made, including radioactivity levels and performance
characteristics. These have been used to inform a complete end-to-end Monte
Carlo simulation that has then been used to calculate the expected performance
of the new instrument, both operating alone and as an anti-coincidence detector
for ZEPLIN-III. The veto device will be capable of rejecting over 65% of
coincident nuclear recoil events from neutron background in the energy range of
interest in ZEPLIN-III. This will reduce the background in ZEPLIN-III from ~0.4
to ~0.14 events per year in the WIMP acceptance region, a significant factor in
the event of a non-zero observation. Furthermore, in addition to providing
valuable diagnostic capabilities, the veto is capable of tagging over 15% for
gamma-ray rejection, all whilst contributing no significant additional
background. In conjunction with the replacement of the internal ZEPLIN-III
photomultiplier array, the new veto is expected to improve significantly the
sensitivity of the ZEPLIN-III instrument to dark matter, allowing spin
independent WIMP-nucleon cross sections below 1E-8 pb to be probed.
-
We present limits on the WIMP-nucleon cross section for inelastic dark matter
derived from the 2008 run of ZEPLIN-III. Cuts, notably on scintillation pulse
shape and scintillation-to-ionisation ratio, give a net exposure of 63 kg.days
in the range 20-80keV nuclear recoil energy, in which 6 events are observed.
Upper limits on signal rate are derived from the maximum empty patch in the
data. Under standard halo assumptions a small region of parameter space
consistent, at 99% CL, with causing the 1.17 ton.year DAMA modulation signal is
allowed at 90% CL: it is in the mass range 45-60 GeV with a minimum CL of 88%,
again derived from the maximum patch. This is the tightest constraint on that
explanation of the DAMA result yet presented using a xenon target.
-
The ZEPLIN-III experiment in the Palmer Underground Laboratory at Boulby uses
a 12kg two-phase xenon time projection chamber to search for the weakly
interacting massive particles (WIMPs) that may account for the dark matter of
our Galaxy. The detector measures both scintillation and ionisation produced by
radiation interacting in the liquid to differentiate between the nuclear
recoils expected from WIMPs and the electron recoil background signals down to
~10keV nuclear recoil energy. An analysis of 847kg.days of data acquired
between February 27th 2008 and May 20th 2008 has excluded a WIMP-nucleon
elastic scattering spin-independent cross-section above 8.1x10(-8)pb at
55GeV/c2 with a 90% confidence limit. It has also demonstrated that the
two-phase xenon technique is capable of better discrimination between electron
and nuclear recoils at low-energy than previously achieved by other xenon-based
experiments.
-
A method is described that allows calibration and assessment of the linearity
of response of an array of photomultiplier tubes. The method does not require
knowledge of the photomultiplier single photoelectron response model and uses
science data directly, thus eliminating the need for dedicated data sets. In
this manner all photomultiplier working conditions (e.g. temperature, external
fields, etc) are exactly matched between calibration and science acquisitions.
This is of particular importance in low background experiments such as
ZEPLIN-III, where methods involving the use of external light sources for
calibration are severely constrained.