The scientific performance of the Planck Low Frequency Instrument (LFI) after
one year of in-orbit operation is presented. We describe the main optical
parameters and discuss photometric calibration, white noise sensitivity, and
noise properties. A preliminary evaluation of the impact of the main systematic
effects is presented. For each of the performance parameters, we outline the
methods used to obtain them from the flight data and provide a comparison with
pre-launch ground assessments, which are essentially confirmed in flight.
-
The Low Frequency Instrument (LFI) on-board the ESA Planck satellite carries
eleven radiometer subsystems, called Radiometer Chain Assemblies (RCAs), each
composed of a pair of pseudo-correlation receivers. We describe the on-ground
calibration campaign performed to qualify the flight model RCAs and to measure
their pre-launch performances. Each RCA was calibrated in a dedicated
flight-like cryogenic environment with the radiometer front-end cooled to 20K
and the back-end at 300K, and with an external input load cooled to 4K. A
matched load simulating a blackbody at different temperatures was placed in
front of the sky horn to derive basic radiometer properties such as noise
temperature, gain, and noise performance, e.g. 1/f noise. The spectral response
of each detector was measured as was their susceptibility to thermal variation.
All eleven LFI RCAs were calibrated. Instrumental parameters measured in these
tests, such as noise temperature, bandwidth, radiometer isolation, and
linearity, provide essential inputs to the Planck-LFI data analysis.
-
The 30 and 44 GHz Back End Modules (BEM) for the Planck Low Frequency
Instrument are broadband receivers (20% relative bandwidth) working at room
temperature. The signals coming from the Front End Module are amplified, band
pass filtered and finally converted to DC by a detector diode. Each receiver
has two identical branches following the differential scheme of the Planck
radiometers. The BEM design is based on MMIC Low Noise Amplifiers using GaAs
P-HEMT devices, microstrip filters and Schottky diode detectors. Their
manufacturing development has included elegant breadboard prototypes and
finally qualification and flight model units. Electrical, mechanical and
environmental tests were carried out for the characterization and verification
of the manufactured BEMs. A description of the 30 and 44 GHz Back End Modules
of Planck-LFI radiometers is given, with details of the tests done to determine
their electrical and environmental performances. The electrical performances of
the 30 and 44 GHz Back End Modules: frequency response, effective bandwidth,
equivalent noise temperature, 1/f noise and linearity are presented.
-
The Planck Low Frequency Instrument (LFI) radiometers have been tested
extensively during several dedicated campaigns. The present paper reports the
principal noise properties of the LFI radiometers.
-
In this paper we discuss the linearity response of the Planck-LFI receivers,
with particular reference to signal compression measured on the 30 and 44 GHz
channels. In the article we discuss the various sources of compression and
present a model that accurately describes data measured during tests performed
with individual radiomeric chains. After discussing test results we present the
best parameter set representing the receiver response and discuss the impact of
non linearity on in-flight calibration, which is shown to be negligible.
-
We give a description of the design, construction and testing of the 30 and
44 GHz Front End Modules (FEMs) for the Low Frequency Instrument (LFI) of the
Planck mission to be launched in 2009. The scientific requirements of the
mission determine the performance parameters to be met by the FEMs, including
their linear polarization characteristics.
The FEM design is that of a differential pseudo-correlation radiometer in
which the signal from the sky is compared with a 4-K blackbody load. The Low
Noise Amplifier (LNA) at the heart of the FEM is based on indium phosphide High
Electron Mobility Transistors (HEMTs). The radiometer incorporates a novel
phase-switch design which gives excellent amplitude and phase match across the
band.
The noise temperature requirements are met within the measurement errors at
the two frequencies. For the most sensitive LNAs, the noise temperature at the
band centre is 3 and 5 times the quantum limit at 30 and 44 GHz respectively.
For some of the FEMs, the noise temperature is still falling as the ambient
temperature is reduced to 20 K. Stability tests of the FEMs, including a
measurement of the 1/f knee frequency, also meet mission requirements.
The 30 and 44 GHz FEMs have met or bettered the mission requirements in all
critical aspects. The most sensitive LNAs have reached new limits of noise
temperature for HEMTs at their band centres. The FEMs have well-defined linear
polarization characteristcs.
-
We give the calibration and scientific performance parameters of the Planck
Low Frequency Instrument (LFI) measured during the ground cryogenic test
campaign. These parameters characterise the instrument response and constitute
our best pre-launch knowledge of the LFI scientific performance. The LFI shows
excellent $1/f$ stability and rejection of instrumental systematic effects;
measured noise performance shows that LFI is the most sensitive instrument of
its kind. The set of measured calibration parameters will be updated during
flight operations through the end of the mission.
-
In this paper we present the Low Frequency Instrument (LFI), designed and
developed as part of the Planck space mission, the ESA program dedicated to
precision imaging of the cosmic microwave background (CMB). Planck-LFI will
observe the full sky in intensity and polarisation in three frequency bands
centred at 30, 44 and 70 GHz, while higher frequencies (100-850 GHz) will be
covered by the HFI instrument. The LFI is an array of microwave radiometers
based on state-of-the-art Indium Phosphide cryogenic HEMT amplifiers
implemented in a differential system using blackbody loads as reference
signals. The front-end is cooled to 20K for optimal sensitivity and the
reference loads are cooled to 4K to minimise low frequency noise. We provide an
overview of the LFI, discuss the leading scientific requirements and describe
the design solutions adopted for the various hardware subsystems. The main
drivers of the radiometric, optical and thermal design are discussed, including
the stringent requirements on sensitivity, stability, and rejection of
systematic effects. Further details on the key instrument units and the results
of ground calibration are provided in a set of companion papers.
-
This paper provides an overview of the Low Frequency Instrument (LFI)
programme within the ESA Planck mission. The LFI instrument has been developed
to produce high precision maps of the microwave sky at frequencies in the range
27-77 GHz, below the peak of the cosmic microwave background (CMB) radiation
spectrum. The scientific goals are described, ranging from fundamental
cosmology to Galactic and extragalactic astrophysics. The instrument design and
development are outlined, together with the model philosophy and testing
strategy. The instrument is presented in the context of the Planck mission. The
LFI approach to ground and inflight calibration is described. We also describe
the LFI ground segment. We present the results of a number of tests
demonstrating the capability of the LFI data processing centre (DPC) to
properly reduce and analyse LFI flight data, from telemetry information to
calibrated and cleaned time ordered data, sky maps at each frequency (in
temperature and polarization), component emission maps (CMB and diffuse
foregrounds), catalogs for various classes of sources (the Early Release
Compact Source Catalogue and the Final Compact Source Catalogue). The
organization of the LFI consortium is briefly presented as well as the role of
the core team in data analysis and scientific exploitation. All tests carried
out on the LFI flight model demonstrate the excellent performance of the
instrument and its various subunits. The data analysis pipeline has been tested
and its main steps verified. In the first three months after launch, the
commissioning, calibration, performance, and verification phases will be
completed, after which Planck will begin its operational life, in which LFI
will have an integral part.
-
The LFI (Low Frequency Instrument) on board the ESA Planck satellite is
constituted by an array of radiometric detectors actively cooled at 20 K in the
30-70 GHz frequency range in the focal plane of the Planck telescope. In this
paper we present an overview of the LFI instrument, with a particular focus on
the radiometer design. The adopted pseudo-correlation scheme uses a software
balancing technique (with a tunable parameter called gain modulation factor)
which is effective in reducing the radiometer susceptibility to amplifier
instabilities also in presence of small non-idealities in the radiometric chain
components, provided that the gain modulation factor is estimated with an
accuracy of the order of 0.2%. These results have been recently confirmed by
experimental laboratory measurements conducted on the LFI prototype radiometers
at 30, 70 and 100 GHz.