The so-called "optimal filter" analysis of a microcalorimeter's x-ray pulses
is statistically optimal only if all pulses have the same shape, regardless of
energy. The shapes of pulses from a nonlinear detector can and do depend on the
pulse energy, however. A pulse-fitting procedure that we call "tangent
filtering" accounts for the energy dependence of the shape and should therefore
achieve superior energy resolution. We take a geometric view of the
pulse-fitting problem and give expressions to predict how much the energy
resolution stands to benefit from such a procedure. We also demonstrate the
method with a case study of K-line fluorescence from several 3d transition
metals. The method improves the resolution from 4.9 eV to 4.2 eV at the Cu
K$\alpha$ line (8.0keV).
-
High-resolution pionic-atom x-ray spectroscopy was performed with an x-ray
spectrometer based on a 240-pixel array of superconducting
transition-edge-sensor (TES) microcalorimeters at the piM1 beam line of the
Paul Scherrer Institute. X-rays emitted by pionic carbon via the 4f->3d
transition and the parallel 4d->3p transition were observed with a
full-width-at-half-maximum energy resolution of 6.8 eV at 6.4 keV. Measured
x-ray energies are consistent with calculated electromagnetic values which
considered the strong-interaction effect assessed via the Seki-Masutani
potential for the 3p energy level, and favor the electronic population of two
filled 1s electrons in the K-shell. Absolute energy calibration with an
uncertainty of 0.1 eV was demonstrated under a high-rate hadron beam condition
of 1.45 MHz. This is the first application of a TES spectrometer to
hadronic-atom x-ray spectroscopy and is an important milestone towards
next-generation high-resolution kaonic-atom x-ray spectroscopy.
-
The analysis of data from x-ray microcalorimeters requires great care; their
excellent intrinsic energy resolution cannot usually be achieved in practice
without a statistically near-optimal pulse analysis and corrections for
important systematic errors. We describe the essential parts of a
pulse-analysis pipeline for data from x-ray microcalorimeters, including steps
taken to reduce systematic gain variation and the unwelcome dependence of
filtered pulse heights on the exact pulse-arrival time. We find these steps
collectively to be essential tools for getting the best results from a
microcalorimeter-based x-ray spectrometer.
-
Transition edge sensor microcalorimeters can measure x-ray and gamma-ray
energies with very high energy resolution and high photon-collection
efficiency. For this technology to reach its full potential in future x-ray
observatories, each sensor must be able to measure hundreds or even thousands
of photon energies per second. Current "optimal filtering" approaches to
achieve the best possible energy resolution work only for photons well isolated
in time, a requirement in direct conflict with the need for high-rate
measurements. We describe a new analysis procedure to allow fitting for the
pulse height of all photons even in the presence of heavy pulse pile-up. In the
limit of isolated pulses, the technique reduces to the standard optimal
filtering with long records. We employ reasonable approximations to the noise
covariance function in order to render multi-pulse fitting computationally
viable even for very long data records. The technique is employed to analyze
x-ray emission spectra at 600 eV and 6 keV at rates up to 250 counts per second
in microcalorimeters having exponential signal decay times of approximately 1.2
ms.
-
We are preparing for an ultra-high resolution x-ray spectroscopy of kaonic
atoms using an x-ray spectrometer based on an array of superconducting
transition-edge-sensor microcalorimeters developed by NIST. The instrument has
excellent energy resolutions of 2 - 3 eV (FWHM) at 6 keV and a large collecting
area of about 20 mm^2. This will open new door to investigate kaon-nucleus
strong interaction and provide new accurate charged-kaon mass value.
-
In this paper we present a new measurement setup, where a transitionedge
sensor detector array is used to detect X-rays in particle induced X-ray
emission measurements with a 2 MeV proton beam. Transition-edge sensors offer
orders of magnitude improvement in energy resolution compared to conventional
silicon or germanium detectors, making it possible to recognize spectral lines
in materials analysis that have previously been impossible to resolve, and to
get chemical information from the elements. Our sensors are cooled to the
operation temperature (65 mK) with a cryogen-free adiabatic demagnetization
refrigerator, which houses a specially designed X-ray snout that has a vacuum
tight window to couple in the radiation. For the best pixel, the measured
instrumental energy resolution was 3.06 eV full width at half maximum at 5.9
keV.We discuss the current status of the project, benefits of transition-edge
sensors when used in particle induced X-ray emission spectroscopy, and the
results from the first measurements.
-
SCUBA-2 is an innovative 10000 pixel bolometer camera operating at
submillimetre wavelengths on the James Clerk Maxwell Telescope (JCMT). The
camera has the capability to carry out wide-field surveys to unprecedented
depths, addressing key questions relating to the origins of galaxies, stars and
planets. With two imaging arrays working simultaneously in the atmospheric
windows at 450 and 850 microns, the vast increase in pixel count means that
SCUBA-2 maps the sky 100-150 times faster than the previous SCUBA instrument.
In this paper we present an overview of the instrument, discuss the physical
characteristics of the superconducting detector arrays, outline the observing
modes and data acquisition, and present the early performance figures on the
telescope. We also showcase the capabilities of the instrument via some early
examples of the science SCUBA-2 has already undertaken. In February 2012,
SCUBA-2 began a series of unique legacy surveys for the JCMT community. These
surveys will take 2.5 years and the results are already providing complementary
data to the shorter wavelength, shallower, larger-area surveys from Herschel.
The SCUBA-2 surveys will also provide a wealth of information for further study
with new facilities such as ALMA, and future telescopes such as CCAT and SPICA.
-
We demonstrate the code-division multiplexed (CDM) readout of eight
transition-edge sensor microcalorimeters. The energy resolution is 3.0 eV (full
width at half-maximum) or better at 5.9 keV, with a best resolution of 2.3 eV
and a mean of 2.6 eV over the seven modulated detectors. The flux-summing CDM
system is described and compared with similar time-division multiplexed (TDM)
readout. We show that the sqrt(Npixels) multiplexing disadvantage associated
with TDM is not present in CDM. This demonstration establishes CDM as both a
simple route to higher performance in existing TDM microcalorimetric
experiments and a long-term approach to reaching higher multiplexing factors.
-
We are developing code-division multiplexing (CDM) systems for
transition-edge sensor arrays with the goal of reaching multiplexing factors in
the hundreds. We report on x-ray measurements made with a four-channel
prototype CDM system that employs a flux-summing architecture, emphasizing
data-analysis issues. We describe an empirical method to determine the
demodulation matrix that minimizes cross-talk. This CDM system achieves energy
resolutions of between 2.3 eV and 3.0 eV FWHM at 5.9 keV.
-
Multiplexers based on the modulation of superconducting quantum interference
devices are now regularly used in multi-kilopixel arrays of superconducting
detectors for astrophysics, cosmology, and materials analysis. Over the next
decade, much larger arrays will be needed. These larger arrays require new
modulation techniques and compact multiplexer elements that fit within each
pixel. We present a new in-focal-plane code-division multiplexer that provides
multiplexing elements with the required scalability. This code-division
multiplexer uses compact lithographic modulation elements that simultaneously
multiplex both signal outputs and superconducting transition-edge sensor (TES)
detector bias voltages. It eliminates the shunt resistor used to voltage bias
TES detectors, greatly reduces power dissipation, allows different dc bias
voltages for each TES, and makes all elements sufficiently compact to fit
inside the detector pixel area. These in-focal-plane code-division multiplexers
can be combined with multi-gigahertz readout based on superconducting
microresonators to scale to even larger arrays.
-
We report on twenty-three clusters detected blindly as Sunyaev-Zel'dovich
(SZ) decrements in a 148 GHz, 455 square-degree map of the southern sky made
with data from the Atacama Cosmology Telescope 2008 observing season. All SZ
detections announced in this work have confirmed optical counterparts. Ten of
the clusters are new discoveries. One newly discovered cluster, ACT-CL
J0102-4915, with a redshift of 0.75 (photometric), has an SZ decrement
comparable to the most massive systems at lower redshifts. Simulations of the
cluster recovery method reproduce the sample purity measured by optical
follow-up. In particular, for clusters detected with a signal-to-noise ratio
greater than six, simulations are consistent with optical follow-up that
demonstrated this subsample is 100% pure. The simulations further imply that
the total sample is 80% complete for clusters with mass in excess of 6x10^14
solar masses referenced to the cluster volume characterized by five hundred
times the critical density. The Compton y -- X-ray luminosity mass comparison
for the eleven best detected clusters visually agrees with both self-similar
and non-adiabatic, simulation-derived scaling laws.
-
We report on extragalactic sources detected in a 455 square-degree map of the
southern sky made with data at a frequency of 148 GHz from the Atacama
Cosmology Telescope 2008 observing season. We provide a catalog of 157 sources
with flux densities spanning two orders of magnitude: from 15 to 1500 mJy.
Comparison to other catalogs shows that 98% of the ACT detections correspond to
sources detected at lower radio frequencies. Three of the sources appear to be
associated with the brightest cluster galaxies of low redshift X-ray selected
galaxy clusters. Estimates of the radio to mm-wave spectral indices and
differential counts of the sources further bolster the hypothesis that they are
nearly all radio sources, and that their emission is not dominated by
re-emission from warm dust. In a bright (>50 mJy) 148 GHz-selected sample with
complete cross-identifications from the Australia Telescope 20 GHz survey, we
observe an average steepening of the spectra between 5, 20, and 148 GHz with
median spectral indices of $\alpha_{\rm 5-20} = -0.07 \pm 0.06$, $\alpha_{\rm
20-148} = -0.39 \pm0.04$, and $\alpha_{\rm 5-148} = -0.20 \pm 0.03$. When the
measured spectral indices are taken into account, the 148 GHz differential
source counts are consistent with previous measurements at 30 GHz in the
context of a source count model dominated by radio sources. Extrapolating with
an appropriately rescaled model for the radio source counts, the Poisson
contribution to the spatial power spectrum from synchrotron-dominated sources
with flux density less than 20 mJy is $C^{\rm Sync} = (2.8 \pm 0.3) \times
10^{-6} \micro\kelvin^2$.
-
The Atacama Cosmology Telescope (ACT) is currently observing the cosmic
microwave background with arcminute resolution at 148 GHz, 218 GHz, and 277
GHz. In this paper, we present ACT's first results. Data have been analyzed
using a maximum-likelihood map-making method which uses B-splines to model and
remove the atmospheric signal. It has been used to make high-precision beam
maps from which we determine the experiment's window functions. This beam
information directly impacts all subsequent analyses of the data. We also used
the method to map a sample of galaxy clusters via the Sunyaev-Zel'dovich (SZ)
effect, and show five clusters previously detected with X-ray or SZ
observations. We provide integrated Compton-y measurements for each cluster. Of
particular interest is our detection of the z = 0.44 component of A3128 and our
current non-detection of the low-redshift part, providing strong evidence that
the further cluster is more massive as suggested by X-ray measurements. This is
a compelling example of the redshift-independent mass selection of the SZ
effect.
-
We present a measurement of the angular power spectrum of the cosmic
microwave background (CMB) radiation observed at 148 GHz. The measurement uses
maps with 1.4' angular resolution made with data from the Atacama Cosmology
Telescope (ACT). The observations cover 228 square degrees of the southern sky,
in a 4.2-degree-wide strip centered on declination 53 degrees South. The CMB at
arcminute angular scales is particularly sensitive to the Silk damping scale,
to the Sunyaev-Zel'dovich (SZ) effect from galaxy clusters, and to emission by
radio sources and dusty galaxies. After masking the 108 brightest point sources
in our maps, we estimate the power spectrum between 600 < \ell < 8000 using the
adaptive multi-taper method to minimize spectral leakage and maximize use of
the full data set. Our absolute calibration is based on observations of Uranus.
To verify the calibration and test the fidelity of our map at large angular
scales, we cross-correlate the ACT map to the WMAP map and recover the WMAP
power spectrum from 250 < ell < 1150. The power beyond the Silk damping tail of
the CMB is consistent with models of the emission from point sources. We
quantify the contribution of SZ clusters to the power spectrum by fitting to a
model normalized at sigma8 = 0.8. We constrain the model's amplitude ASZ < 1.63
(95% CL). If interpreted as a measurement of sigma8, this implies sigma8^SZ <
0.86 (95% CL) given our SZ model. A fit of ACT and WMAP five-year data jointly
to a 6-parameter LCDM model plus terms for point sources and the SZ effect is
consistent with these results.
-
The Atacama Cosmology Telescope was designed to measure small-scale
anisotropies in the Cosmic Microwave Background and detect galaxy clusters
through the Sunyaev-Zel'dovich effect. The instrument is located on Cerro Toco
in the Atacama Desert, at an altitude of 5190 meters. A six-meter off-axis
Gregorian telescope feeds a new type of cryogenic receiver, the Millimeter
Bolometer Array Camera. The receiver features three 1000-element arrays of
transition-edge sensor bolometers for observations at 148 GHz, 218 GHz, and 277
GHz. Each detector array is fed by free space mm-wave optics. Each frequency
band has a field of view of approximately 22' x 26'. The telescope was
commissioned in 2007 and has completed its third year of operations. We discuss
the major components of the telescope, camera, and related systems, and
summarize the instrument performance.
-
We report on the design, first observing season, and analysis of data from a
new prototype millimeter-wave interferometer, MINT. MINT consists of four 145
GHz SIS mixers operating in double-sideband mode in a compact heterogeneous
configuration. The signal band is subdivided by a monolithic channelizer, after
which the correlations between antennas are performed digitally. The typical
receiver sensitivity in a 2 GHz band is 1.4 mK sqrt(s). MINT observed the
cosmic microwave background (CMB) from the Chilean Altiplano. The site has a
median nighttime atmospheric temperature of 9 K at zenith (exclusive of the
CMB). Observations of Mars, Jupiter, and a telescope-mounted calibration source
establish the system's phase and magnitude stability. MINT is the first
CMB-dedicated interferometer to operate above 50 GHz. The same type of system
can be used to probe the Sunyaev-Zel'dovich effect in galaxy clusters near the
SZ null at 217 GHz. We present an analysis of sideband-separated, digitally
sampled data recorded by the array. Based on 215 hours of data taken in late
2001, we set an upper limit on the CMB anisotropy in a band of width Delta
ell=700 around ell=1540 of delta T < 105 microK (95% conf). Increased
sensitivity can be achieved with more integration time, greater bandwidth, and
more elements.